![Tektronix 516 series Instruction Manual Download Page 15](http://html1.mh-extra.com/html/tektronix/516-series/516-series_instruction-manual_1078919015.webp)
O perating Instructions — Type 5 1 6
Selecting the Trigger Source
CAL. OUT
D isplay several cycles of C a lib ra to r w aveform on crt screen.
Hold probe barrel and loosen locking sleeve.
Hold probe
base and adjust probe barrel fo r fla t-to p p e d square waves.
Hold probe barrel and carefully tigh ten locking sleeve.
Fig. 3-2. Probe compensation adjustment.
An adjustable capacitor in the probe body compensates
for variations in the input capacitance from one instrument
to another or between two input connectors of one instru
ment. To insure the accuracy of pulse and transient measure
ments, this adjustment should be checked frequently. Fig.
3-2 illustrates the method of adjusting the compensation of
the probe.
TRIGGERED OPERATION
For most uses of your oscilloscope, you will need to get a
stable display of some recurrent waveform. In order to ob
tain a stable display, it is necessary to start the horizontal
sweep consistently at the same time relative to recurring cy
cles of the input waveform. The sweep therefore must be
triggered by the input waveform itself or by some waveform
which bears a fixed time relationship to the input waveform.
The following instructions tell you how to select and use the
proper triggering signal for various applications of the oscil
loscope.
For most applications the sweep can be triggered by the
input waveform. The only requirement is that the input
signal be large enough to provide at least one-fifth of a
major graticule division of deflection on the screen at the
sensitivity level for which the VOLTS/DIV. control is set. To
obtain triggering of the sweep from the input waveform, set
the black TRIGGER SELECTOR control to either the + INT.
or the —INT. position (the significance of the + and — will
be explained in a later paragraph).
For dual-trace operation, it is usually advantageous to
trigger the sweep with some external trigger. In this way,
the true phase relationship of the two waveforms can be
shown and triggering is obtained more easily. In order to
obtain a stable display, it is necessary that the external trig
gering waveform bear a fixed time relationship to the input
waveforms. The external triggering signal must be at least
0.2 volt in amplitude to assure proper triggering. To use an
external waveform for triggering the horizontal sweep, con
nect the signal to the TRIGGER INPUT connector and set the
black TRIGGER SELECTOR control to the +EXT. or —EXT.
position. External triggering may also be used with single
trace operation where desired.
When you are observing a waveform which bears a fixed
time relationship to the ac line frequency, you may wish to
trigger the sweep from the line-frequency waveform. To do
this, place the black TRIGGER SELECTOR control in the
+LINE or —LINE position.
When the black TRIGGER SELECTOR control is set to one
of the
+
positions (+INT., +EXT., -+ LINE), the sweep is trig
gered on a positive slope of the triggering signal. When the
control is set to one of the — positions (—INT., —EXT.,
—LINE), the sweep is triggered on a negative slope of the
triggering signal. In most cases, selection of the triggering
slope is not critical.
Selecting the Triggering Mode (HF SYNC,
AUTO., AC, DC)
The most useful triggering mode is the automatic mode,
obtained by setting the red TRIGGER SELECTOR control
to the AUTO, position. The automatic mode of triggering
provides a stable display with virtually any triggering signal
from about 50 Flz to 2 MFlz in frequency. No adjustment
of the STABILITY or TRIGGERING LEVEL controls is necessary
since these two functions are automatically set to their op
timum values internally. (In the other triggering modes, it
is necessary to set one or both of these controls to obtain
a stable presentation.) In addition, the automatic mode of
triggering provides a free-running trace on the screen in
the absence of any triggering signal. This trace may be
used, under certain conditions, as a base line from which to
make dc measurements. It also gives a constant indication
of the readiness of the oscilloscope to accept a triggering
signal for the display of a stable waveform.
In the ac triggering mode, obtained by setting the red
TRIGGER SELECTOR control in the AC position, suitable
triggering can be obtained with signals from about 16 Flz
to 5 MFfz. When the ac triggering mode is used, it is usually
d
3-6
Summary of Contents for 516 series
Page 5: ... Type 516 ...
Page 19: ...NOTES Type 516 ...
Page 39: ...NOTES Type 516 ...
Page 54: ...Type 516 SECTION 7 MECHANICAL PARTS LIST 7 1 ...
Page 55: ... Mechanical Parts List Type 516 FRONT 7 2 ...
Page 61: ...Mechanical Parts List Type 516 REAR 7 8 ...
Page 67: ...Mechanical Part List Type 516 LEFTSIDE 7 14 ...
Page 73: ...Mechanical Parts List Type 516 TOP 7 20 ...
Page 78: ...POWER VERTICAL AMPLIFIER CHASSIS Mechanical Parts List Type 516 7 25 ...
Page 107: ...INPUT A INPUT ft T R IG G E R IN P U T C A L O U T A TYPE S I6 OSCILLOSCOPE ...
Page 111: ... jY P O S C ILLO S C O P E F ...
Page 114: ...T 1 1 TIM IN G I69 e SWITCH ...
Page 116: ...C I R C U I T M U M tp S 2 2 5 T H R U 3 8 9 H O R IZ O N T A L A M P L IF IE R ...
Page 117: ... ...
Page 124: ...33 K UNBLANKING PULSE C IR C U IT N U M tP S 6 0 0 TH g U 8 6 9 C R T C IR C U IT TP o oA ...