![Tektronix 516 series Instruction Manual Download Page 24](http://html1.mh-extra.com/html/tektronix/516-series/516-series_instruction-manual_1078919024.webp)
Circuit Description — Type 5 1 6
FROM
TO
TIME-BASE
GENERATOR
Fig. 4 -4 . Time-Base Trigger block diagram .
Trigger Input Amplifier
The
input to the
Trigger Input Amplifier, V24,
may
be se
lected from any one of three sources by means of the black
TRIGGER SELECTOR control, SW10A. When SW10A is in
one of the INT. positions, the signal is obtained from the
Trigger Pickoff circuit in the Vertical Amplifier. When SW10A
is in one of the EXT. positions, the signal may be obtained
from an external source through the TRIGGER INPUT con
nector on the front panel. When SW10A is in one of the
LINE positions, the signal is obtained from one of the 6.3-
volt secondary windings of the line transformer.
The + and — positions of SW10A provide a means of in
verting or not inverting, as desired, the triggering signal in
the Trigger Input Amplifier. This is done so that the negative
spike at the output of the Time-Base Trigger can be made
to occur during either a positive-going or a negative-going
portion of the triggering signal. (The negative spike occurs
only when there is a negative-going signal at the input to
the Trigger Multivibrator.)
When SW10A is in any of the —positions, the incoming
triggering signal is applied to the grid of V24A, and V24 is
a cathode-coupled amplifier. With this configuration, the
signal at the plate of V24B (output of the Trigger Input Am
plifier) is in phase with the incoming triggering signal. When
SW10A is in any of the + positions, the incoming triggering
signal is applied to the grid of V24B, and V24B is a plate-
loaded amplifier. In this case, the signal at the plate of
V24B is 180° out of phase with the incoming triggering
signal.
The TRIGGERING LEVEL control, R17, varies the average
dc level at the plate of V24B from about + 60 volts to +100
volts. This is true whether SW10A is in a + position or a
— position. The minimum level of + 60 volts represents the
point where V24B is taken into saturation by a sufficiently
positive setting of R17 when SW10A is in a — position or
by a sufficiently negative setting of R17 when SW10A is in
a + position. The maximum level of +100 volts represents
the point where V24B is taken into cutoff by a sufficiently
negative setting of R17 when SW10A is in a — position or
by a sufficiently positive setting of R17 when SW10A is in
a + position. As will be seen later, the voltage at the plate
of V24B must pass through the approximate center of this
range (about + 80 volts) in order to cause the Trigger Multi
vibrator to change states.
For small triggering signals, R17 is set such that the aver
age dc level at the plate of V24B is close to the center of
its range so that the small signal, as amplified by V24, is
sufficient to carry the plate voltage through the +80-volt
point. When a large triggering signal is applied and it is
desired to trigger on an extreme positive or negative point
of it, R17 is set such that V24B is well into saturation, or cut
off (depending upon whether triggering is desired at a more
negative or more positive point on the signal and on a nega
tive-going or positive-going slope). In this case, the trig
gering signal must be sufficient to overcome the saturation
or cutoff of V24B and produce an additional 20 volts of
swing at the plate of V24B in order to cause the Trigger
Multivibrator to switch states.
It should be noted that the voltages given in the foregoing
discussion are typical nominals only and will vary somewhat
from instrument to instrument and with time.
Trigger Multivibrator
The Trigger Multivibrator is a two-state Schmitt circuit.
When the voltage at its input grid (grid of V45A) is above
a certain critical level (neglecting hysteresis), the Trigger
Multivibrator is in one state, with V45A conducting and
V45B cut off. When the Trigger Multivibrator is in this state,
the voltage as its output (plate of V45B) is at 300 volts.
When the voltage at the input grid is below the critical level
(still neglecting hysteresis), the Trigger Multivibrator is in the
other state, with V45A cut off and V45B conducting. When
the Trigger Multivibrator is in this second state, the voltage
at its output is about +290 volts. The transition from one
4-5
Summary of Contents for 516 series
Page 5: ... Type 516 ...
Page 19: ...NOTES Type 516 ...
Page 39: ...NOTES Type 516 ...
Page 54: ...Type 516 SECTION 7 MECHANICAL PARTS LIST 7 1 ...
Page 55: ... Mechanical Parts List Type 516 FRONT 7 2 ...
Page 61: ...Mechanical Parts List Type 516 REAR 7 8 ...
Page 67: ...Mechanical Part List Type 516 LEFTSIDE 7 14 ...
Page 73: ...Mechanical Parts List Type 516 TOP 7 20 ...
Page 78: ...POWER VERTICAL AMPLIFIER CHASSIS Mechanical Parts List Type 516 7 25 ...
Page 107: ...INPUT A INPUT ft T R IG G E R IN P U T C A L O U T A TYPE S I6 OSCILLOSCOPE ...
Page 111: ... jY P O S C ILLO S C O P E F ...
Page 114: ...T 1 1 TIM IN G I69 e SWITCH ...
Page 116: ...C I R C U I T M U M tp S 2 2 5 T H R U 3 8 9 H O R IZ O N T A L A M P L IF IE R ...
Page 117: ... ...
Page 124: ...33 K UNBLANKING PULSE C IR C U IT N U M tP S 6 0 0 TH g U 8 6 9 C R T C IR C U IT TP o oA ...