Operating Instructions— 442 Service
NORTH AMERICAN
120 VOLT/60 Hz
RATED 15 AMPERE
(12 AMP MAX.; NEC)
UNIVERSAL EURO
220 VOLT/50 Hz
RATED 16 AMPERE
240 VOLT/60 Hz
RATED 15 AMPERE
(12 AMP MAX.; NEC)
U.K.
240 VOLT/50 Hz
RATED 13 AMPERE
AUSTRALIAN
240 VOLT/50 Hz
RATED 10 AMPERE
( 2 3 8 9 - 7 6 ) 2 7 5 7 - 6 6
AC LINE VOLTAGE
HI/LO
SELECTION (120 V/240 V)
/
\
/
Figure 2-2. Power switch locations on bottom of the 442.
Table 2-2
AC REGULATING RANGES
HI/LO
Switch
Position
120 V/240 V Setting
120 V
240 V
HI
110 to 132 V rms
220 to 250 V rms
LO
92 to 110 V rms
202 to 240 V rms
Fuse Information
The 442 power input fuse is located in a fuse holder on
the back panel of the 442. See Figure 2-3.
Figure 2-1. Commonly used power plugs.
Set the AC Line Voltage Selection syvitchtothe nominal
line voltage available (see Table 2-2). The fuse value must
be changed when a different Line Voltage range is
selected (see Table 2-3).
Setthe HI/LO Regulating Range switch (see Figure 2-2)
so the expected line-voltage fluctuations remain within
the range selected (see Table 2-2).
Table 2-3
POWER INPUT FUSE VALUES
Applied Voltage
F700 Value
120 V ac
0.75 A Fast Blow
240 V ac
0.30 A Slow Blow
2 -2
REV A MAY 1980
Summary of Contents for 442
Page 9: ...442 Service 442 Oscilloscope viii REV A MAY 1980 ...
Page 113: ...Figure 7 4 442 Oscilloscope block diagram REV A JUN 1980 2 3 7 4 2 7 A ...
Page 117: ...A8 VERTICAL BOARD Figure 7 6 Vertical circuit board location ...
Page 118: ......
Page 120: ......
Page 122: ......
Page 123: ...Figure 7 8 interface circuit board location ...
Page 124: ......
Page 126: ......
Page 130: ...A ll TRIGGER SWITCH BOARD 2 3 7 4 3 6 A Figure 7 10 Trigger Switch circuit board location ...
Page 131: ...2 3 7 4 S 4 R ev jun VRSO V E R T IC A L OUT PUT CRT t PROBE ADJ 3 ...
Page 133: ......
Page 137: ......
Page 139: ......
Page 140: ...4 4 Z ...
Page 142: ......
Page 148: ...R EV K JUN 9SO ...
Page 153: ... 442 ...
Page 154: ...442 REV A DEC 1978 ...
Page 164: ......