Recalibration
—
Type
575
15. Checking
Vertical
Collector
ma/div
Switch
Set
controls
as
follows:
Horizontal
20
COLLECTOR
VOLTS
Vertical
.01
MA/STEP
STEP
SELECTOR
.01
MA/STEP
Base
Step
Generator
REPETITIVE
POLARITY
—
PEAK
VOLTS
0
DISSIPATION
LIMITING
RESISTOR
0
Short
binding posts
C
and
B on
right
side
of
test
panel
together
and
switch
Transistor
Selector
Switch
to
TRANSISTOR
B.
Note
a
display
of
one
dot
per
division
±1%
minus
the
percent
of
error
recorded
in
step
1
(Table
5-1)
for
the
.01
MA
PER
STEP
position
of
the
STEP
SELECTOR
switch.
Do this
procedure
for
each
of
the
corresponding
positions
of
the
CURRENT
OR
VOLTAGE
PER
DIVISION
section
of
the
Vertical
switch
and
the
MA
PER
STEP
section
of
the
STEP
SELECTOR
switch.
The
500 and
1000
COLLECTOR
MA
positions
of
the
Vertical
switch
should
be
checked
with
the STEP
SELECTOR
at
50
and
100
MA
PER
STEP
respectively,
and with
the
0.1
X
button
depressed.
5-5
Summary of Contents for 175
Page 4: ...VWTICM OINltAlO TRANSISTOR CUM TRACfR UB l COUlClO r hokizoniai VCKIJ O lUnK Hix M Type 575 z ...
Page 6: ...NOTES ...
Page 26: ...NOTES ...
Page 48: ...NOTES ...
Page 54: ...Recalibration Type 575 C706 Fig 5 3 Left Side View 5 6 ...
Page 56: ...NOTES ...
Page 100: ...INPUT AMPLIF1E ...
Page 104: ...TYPE 575 TRANSISTOR CURVE TRACER ...
Page 108: ...OSCILLATOR 4OOV UNREG TYPE 575 TRAN5I5TOR CURVE TRACER AB ...
Page 115: ... 300 ...
Page 119: ...Type 175 Tent S N 570 PARTS LIST CORRECTION Change to Q233 151 0137 00 2N21U8 MIO 298 1165 ...