Operator Tutorial
1503C MTDR User Manual
2–21
O
F
F
O
F
F
O
F
F
O
N
ac
0.00 ft
Figure 2–32: Display with Pulse Turned Off
Note that there is no pulse visible on the display. The 1503C is now functioning as
a listening device, much like a non-triggered oscilloscope. Uses include monitoring
the cable for pulses.
11. To turn the pulse back on, enter the Acquisition Control Menu again, scroll to
Pulse is: Off and press MENU to turn the pulse back on. Repeatedly press
MENU until the instrument returns to normal operation.
1. Set the 1503C front-panel controls to:
CABLE
Attach 10-ft cable
IMPEDANCE
93
NOISE FILTER
1 avg
VERT SCALE
0.00 dB (default)
DIST/DIV
1 ft/div (0.25 m)
Vp
.84
PULSE WIDTH
2 ns
2. Pull POWER on.
3. Press MENU to access the Main Menu.
4. Using the
n
o
POSITION control, scroll down to Setup Menu.
5. Press MENU to accept this selection.
6. Scroll down to Acquisition Control Menu.
7. Press MENU to accept this selection.
8. Scroll down to Single Sweep is: Off.
9. Press MENU to toggle this selection. It should now read Single Sweep is: On.
10. Press MENU repeatedly until the instrument returns to normal operation. The
waveform on the display is the familiar test cable.
Single Sweep