DMM4040/4050
Technical Reference Manual
Description of IEC 61010 Measurement Categories
The IEC 61010 safety standard defines four Overvoltage (Installation) Categories (CAT I
to CAT IV) based on the magnitude of danger from transient impulses as shown in Figure
1-1.
cat_levels.eps
Figure 1-1. IEC 61010 Measurement Category (CAT) Levels
The IEC 61010 Measurement CAT level indicates the level of protection the instrument
provides against impulse withstand voltage.
CAT I
equipment is designed to protect against transients from high-voltage, low-energy
sources, such as electronic circuits or a copy machine.
CAT II
equipment is designed to protect against transients from energy-consuming
equipment supplied from the fixed installation, such as TVs, PCs, portable tools, and
other household appliances.
CAT III
equipment is designed to protect against transients in equipment in fixed
equipment installations, such as distribution panels, feeders and short branch circuits, and
lighting systems in large buildings.
CAT IV
equipment is designed to protect against transients from the primary supply
level, such as an electricity meter or an overhead or underground utility service.
Organization of the Calibration Manual
This calibration manual is divided into the following chapters:
1-8
Summary of Contents for DMM4040
Page 1: ...DMM4040 and DMM4050 Digital Multimeter Technical Reference 077 0362 01...
Page 6: ...DMM4040 and DMM4050 Technical Reference iv...
Page 8: ...DMM4040 and DMM4050 Technical Reference vi...
Page 10: ...DMM4040 and DMM4050 Technical Reference viii...
Page 34: ...DMM4040 4050 Technical Reference Manual 2 2...
Page 42: ...DMM4040 4050 Technical Reference Manual 3 2...
Page 75: ...Appendices Appendix Title Page A Verification Forms A 1 B Example Adjustment Program B 1...
Page 76: ...DMM4040 4050 Technical Referencel...