Performance Test and Calibration
Performance Tests
3
Table 3-12. DMM4050 Capacitance Verification Steps
Nominal Output
Range
1-year Test Limits
High (F)
Low (F)
0
η
F
1.0
η
F
25.0 pF
0
η
F
1.0
η
F
1.0
η
F
1.045
η
F
955.0 pF
10.0
η
F
10.0
η
F
10.15
η
F
9.85
η
F
100.0
η
F
100.0
η
F
101.5
η
F
98.5
η
F
1.0
µ
F
1.0
µ
F
1.015
µ
F
985.0
η
F
10.0
µ
F
10.0
µ
F
10.15
µ
F
9.85
µ
F
100.0
µ
F
100.0
µ
F
101.5
µ
F
98.5
µ
F
1.0 mF
1.0 mF
1.015 mF
985.0
µ
F
10.0 mF
10.0 mF
10.15 mF
9.85 mF
100.0 mF
100.0 mF
104.2 mF
95.8 mF
3-21