System Installation

12 • IL380 Tek-CARE

®

 NC150 & NC200 Manual

Copyright © TekTone Sound & Signal Mfg., Inc. All Rights Reserved.

System Installation

System Checkout and Testing

IR151B, IR152B, IR155B Patient Stations: Test patient stations one at a time. Initiate 

a call on each patient station: Press the button on the end of the call cord for IR151B 

and IR152B Patient Stations (test both circuits on IR152B stations). Pull the call cord 

on the IR155B.

Check for operation of the following signals: 

• The 

call

 

placed

 

lamp on the patient station should be illuminated. 

•  The LI381, LI382 or LI382LED Corridor Light near the room entrance should 

be illuminated. 

•  On the NC150N or NC200N Master Station, the call lamp should be illuminated 

and the associated station selector lamp (marked to identify the calling station) 

should be illuminated. A slowly repeating audible call tone should be heard. 

The normal call tone may be silenced by the 

tone

 

off

 

button, which should be 

illuminated to indicate tone silencing.

•  On all LI150B Duty Stations, the 

call

 

placed

 lamp should be illuminated, and 

a slowly repeating audible call tone should be heard. The normal call tone may 

be silenced by the 

tone

 

off

 switch.

Reset the call at the patient station: Press the 

reset

 button on IR151B and IR152B 

Stations. Push the 

reset

 

switch to the left on IR155B Stations. All signals should 

be canceled.
Test the intercom function:

•  Initiate a call on each station as previously described.

•  On the master station, select the calling station by pressing the illuminated station 

selector button and check for the following:

1. The 

station

 

selector

 lamp should be canceled.

2. The 

call

 

lamp should turn off.

3.  The audible call tone should be canceled.

4.  The remote station’s 

in

 

use

 

LED should be illuminated.

•  Press the 

call

 button and the 

talk

 button on the master station. An electronic 

call tone should be heard at the master station and at the patient station.

•  Press the 

talk

 button and speak in a normal voice. Release the button to hear 

the reply.

•  At the patient station, the 

call placed

 lamp should turn off, and the 

in use

 LED 

should be illuminated.

•  When voice communication is heard, reply in a normal voice.

•  Return all controls to the normal position.

On each IR150B Staff Station, test operation the same as with the IR151B Patient 

Station, except place a call using the 

call

 button instead of a call cord.

On each SF155B Emergency or SF156B Code station (one at a time), slide the call 

switch down, and then check the following signals:

• The 

call

 

placed

 

lamp should be flashing.

•  The LI381, LI382 or LI382LED Corridor Light near the room entrance should 

be flashing.

•  On the master station, the 

call

 

lamp should be flashing, the associated station 

selector lamp (marked to identify calling station) should be flashing, and an 

intermittent audible call tone should be heard. The 

tone

 

off

 

button must not 

cancel the emergency code call tone.

On all LI150B Duty Stations, check the following signals:

• The 

call

 

placed

 

lamp should be flashing.

•  The rapidly pulsing audible call tone should be heard. The 

tone

 

off

 switch must 

not cancel the emergency call tone. 

Summary of Contents for Tek-CARE NC150N

Page 1: ...ation to staff and patient stations as well as emergency signaling and call indication from remote locations Emergency calls are indicated by a rapid pulsing of all audible and visual signals associat...

Page 2: ...should not be construed as a commitment by TekTone Sound Signal Mfg Inc TekTone Sound Signal Mfg Inc assumes no responsibility or liability for any errors or inaccuracies that may appear in this docum...

Page 3: ...ons 15 LI381 Corridor and LI382 Corridor Zone Lights 16 Replacement Parts 16 Illustrations Figure 1 NC150N and NC200N Control Locations 2 Figure 2 Duty Staff Patient Emergency Code Station Control Loc...

Page 4: ...to protect the equipment from electrostatic discharge ESD effects However since the cause is not in the equipment but in the environment further measures are required of the installer and the user to...

Page 5: ...tation s reset lamp will illuminate To speak to the caller press the talk button while speaking and release to listen When the conversation is complete press the associated station selector button to...

Page 6: ...from associated remote when illuminated Indicates emergency call from remote when flashing 6 Lamp Test Button ST005D Illuminates all lamps on panel when pressed 7 Telephone Handset TA150N option Lift...

Page 7: ...at row The remote stations reset lamps will illuminate Group call may be selected for only one row at a time Press the talk button while speaking NOTE Group call serves as a paging function only No re...

Page 8: ...rd is connected to patient station 7 In Use LED Indicates station is in use when illuminated 8 Reset Switch Cancels the call when pushed to the left 9 Call Cord Places a call when pulled 10 Call Place...

Page 9: ...cord is replaced in the receptacle Cancel a Call Press the reset button The call placed lamp will go off Improper Operation If the patient station does not operate as described contact qualified serv...

Page 10: ...teady illumination of the corridor light that is associated with the calling station Improper Operation If corridor light does not operate as described contact qualified service personnel There are no...

Page 11: ...cur LI150B Duty Stations Locate duty stations as needed and where convenient for operation Location should provide for unobstructed visibility of the call indicator LI381 Corridor Lights Locate corrid...

Page 12: ...R150B IR151B IR152B and IR155B Station in the system Shielded cable must be used 1cond 22selectiveforeachSF155BEmergencyStationorSF156BCodeStation not used in conjunction with a patient station 4 cond...

Page 13: ...6 4 5 2 5 Minimum clearance from live parts on the station to dead metal parts to be not less than 0 5 SF155B Emergency and SF156B Code Stations Install single gang ring or single gang ring double gan...

Page 14: ...sformer primary to power source until entire installation is completed and checked for shorts and grounds Install transformer connection box as shown in Figure 7 Transformer box must be UL approved Mi...

Page 15: ...VA Transformer Connect wires as shown in Figure 16 Do not connect transformer primary to power source until entire installation is completed and has been checked for shorts and grounds Connections Ch...

Page 16: ...ed Onthemasterstation selectthecallingstationbypressingtheilluminatedstation selector button and check for the following 1 The station selector lamp should be canceled 2 The call lamp should turn off...

Page 17: ...it out of selector lamp switch knob Identify filter by room number or other means using dry transfer lettering labels or other similar methods Be sure identified filter is replaced in knob so that le...

Page 18: ...ishardwired locate the associated breaker and turn it off If a means cannot be established to remove power to the SS100 then contact qualified service personnel and do not proceed further The SS106 po...

Page 19: ...estore power to the system For repair or replacement of any other parts contact qualified service personnel A list of replacement parts and numbers appears in the Replacement Parts section IR151B IR15...

Page 20: ...red bulb cover Be sure to place bulb cover over the same bulb from which it was removed Totestlight placecallattheassociatedstation Iflightstilldoesnotwork repeat steps above until corridor lamp and o...

Page 21: ...C 068 amps LI382LED Corridor Zone Light 24 VDC 08 amps PK800A Secondary Power Supply 24 VAC 60 Hz Wiring Notes 1 Selector switch shown is representative of both NC150N and NC200N Master Stations 2 Ter...

Page 22: ...y intercom wiring and it should not be ground at multiple locations MASTER STATION PK151N PK151A PK152 POWER CONTROL UNIT 6 18 POINT PAIRS TRUNK 4 18 SS106 TRANSFORMER 2 18 IR STATION SERIES SF STATIO...

Page 23: ...at multiple locations Figure 11 Block Wiring Diagram with Zone Lamps TRUNK CABLE POINT PAIR SYSTEM ZONE LAMP TO MASTER OVERHEAD VIEW OF HOSPITAL CORRIDOR IR STATION SERIES SF STATION SERIES SF STATION...

Page 24: ...rminal identifications shown are not in actual order 4 Use shielded cable if not in a metal conduit If shielded cable is used NO 22 NO 18 NO 18 LI382 or LI382LED SF151N SF153N G T2 P SF155B LI381 LI38...

Page 25: ...0 VA minimum 2 Terminal identifications shown are not in actual order 4 Use shielded cable if not in a metal conduit If shielded cable is used shield wire replaces No 22 No 18 No 18 LI382 LI382LED T2...

Page 26: ...GREEN L BLUE N U Combined Configuration N O N C Smoke Detector Smoke Detector SWITCH MODULE U N J E TALK LISTEN SELECTOR MODULE TERMINAL MODULE C D P P1 L L X L L X L L X L L X X X X X IR150B 151B 15...

Page 27: ...BLUE N U Combined Configuration NOTES SWITCH MODULE U N J E TALK LISTEN SELECTOR MODULE TERMINAL MODULE C D P P1 L L X L L X L L X L L X X X X X IR150N 151N 152N 153N 155N P F V Q L G P L Y C X R IR15...

Page 28: ...upply PK151A PK152 Control unit Fuse Assy Mount within 2 feet of each other Note T1 and T2 inputs when interconnecting PK800A Secondary Power Supply Do not connect SS106 Transformer to PK151A PK152 Gr...

Page 29: ...IL380 Tek CARE NC150 NC200 Manual 25 Copyright TekTone Sound Signal Mfg Inc All Rights Reserved Figure 17 SK151N Switch Board PCB Assembly...

Page 30: ...low Green Blue on station P Not used Q Q L V 7 pin connector Purple F NOTE Both jumpers on back of station need to be in up position DRAWING IL380 IL381 SF337C WIRING REV1 091603 1 EMERGENCY Note Both...

Page 31: ...OTE Both jumpers on back of station need to be in up position Brown Red Orange Yellow Green Blue on station P Not used Q Q L V 7 pin connector Purple F NOTE Both jumpers on back of station need to be...

Page 32: ...DULE J D E E1 L L L L N N P P CONTROL ENCLOSURE MASTER STATION NC110N SELECTOR MODULE TERMINAL MODULE D J E N PK151A PK152 L L L TO SELECTIVE LAMP WIRE ON NC150N NC200N NC150N NC200N TO PK151A PK152 N...

Reviews: