System Installation
12 • IL380 Tek-CARE
®
NC150 & NC200 Manual
Copyright © TekTone Sound & Signal Mfg., Inc. All Rights Reserved.
System Installation
System Checkout and Testing
IR151B, IR152B, IR155B Patient Stations: Test patient stations one at a time. Initiate
a call on each patient station: Press the button on the end of the call cord for IR151B
and IR152B Patient Stations (test both circuits on IR152B stations). Pull the call cord
on the IR155B.
Check for operation of the following signals:
• The
call
placed
lamp on the patient station should be illuminated.
• The LI381, LI382 or LI382LED Corridor Light near the room entrance should
be illuminated.
• On the NC150N or NC200N Master Station, the call lamp should be illuminated
and the associated station selector lamp (marked to identify the calling station)
should be illuminated. A slowly repeating audible call tone should be heard.
The normal call tone may be silenced by the
tone
off
button, which should be
illuminated to indicate tone silencing.
• On all LI150B Duty Stations, the
call
placed
lamp should be illuminated, and
a slowly repeating audible call tone should be heard. The normal call tone may
be silenced by the
tone
off
switch.
Reset the call at the patient station: Press the
reset
button on IR151B and IR152B
Stations. Push the
reset
switch to the left on IR155B Stations. All signals should
be canceled.
Test the intercom function:
• Initiate a call on each station as previously described.
• On the master station, select the calling station by pressing the illuminated station
selector button and check for the following:
1. The
station
selector
lamp should be canceled.
2. The
call
lamp should turn off.
3. The audible call tone should be canceled.
4. The remote station’s
in
use
LED should be illuminated.
• Press the
call
button and the
talk
button on the master station. An electronic
call tone should be heard at the master station and at the patient station.
• Press the
talk
button and speak in a normal voice. Release the button to hear
the reply.
• At the patient station, the
call placed
lamp should turn off, and the
in use
LED
should be illuminated.
• When voice communication is heard, reply in a normal voice.
• Return all controls to the normal position.
On each IR150B Staff Station, test operation the same as with the IR151B Patient
Station, except place a call using the
call
button instead of a call cord.
On each SF155B Emergency or SF156B Code station (one at a time), slide the call
switch down, and then check the following signals:
• The
call
placed
lamp should be flashing.
• The LI381, LI382 or LI382LED Corridor Light near the room entrance should
be flashing.
• On the master station, the
call
lamp should be flashing, the associated station
selector lamp (marked to identify calling station) should be flashing, and an
intermittent audible call tone should be heard. The
tone
off
button must not
cancel the emergency code call tone.
On all LI150B Duty Stations, check the following signals:
• The
call
placed
lamp should be flashing.
• The rapidly pulsing audible call tone should be heard. The
tone
off
switch must
not cancel the emergency call tone.