System Installation
IL380 Tek-CARE
®
NC150 & NC200 Manual • 11
Copyright © TekTone Sound & Signal Mfg., Inc. All Rights Reserved.
System Installation
Wire Connections
Use crimp-style connectors for all wire connections. Do not use wire nuts.
NC150N, NC200N Master Stations: No internal wiring is necessary for the NC150N
and NC200N Master Stations.
IR150B, IR151B, IR152B, IR155B Staff and Patient Stations: Connect wires as shown
in
Figure 12
.
SF155B Emergency Stations: Connect wires as shown in
Figure 12
.
SF156B Code Stations: Connect wires as shown in
Figure 15
.
LI150B Duty Stations: Connect wires as shown in
Figure 12
.
LI381 Corridor Lights, LI382 and LI382LED Corridor Zone Lights: Connect wires
as shown in
Figure 12
.
PK151A, PK152 Power & Control Unit: Connect wires as shown in
Figure 12
. Also
connect secondary from the SS106 Transformer (24 VAC, 30 VA connections) to
the PK151A or PK152 as shown in
Figure 12
. Do not connect transformer primary
to power source until entire installation is completed and has been checked for
shorts and grounds.
PK800A Secondary Power Supply: Connect wires as shown in
Figure 16
.
SS100 24 VAC 100 VA Transformer: Connect wires as shown in
Figure 16
. Do not
connect transformer primary to power source until entire installation is completed
and has been checked for shorts and grounds.
Connections Checkout
Recheck all connections to equipment. If all wires and connections are satisfactory,
connect primary coil of SS106 Transformer to source of 117 VAC 60 Hz (40 watts max.)
and operation of system can be checked according to
System Test Instructions
next
in this section.
System Test Instructions
Before proceeding with a system test, set all stations to normal conditions as follows:
NC150N and NC200N Master Stations: All push buttons should be in the
out
position.
A depressed button may be released by pressing the button and then releasing it.
IR150B Staff Station: If the
call
placed
lamp is illuminated, press the
reset
button
to reset it.
IR151B, IR152B Patient Stations: Insert a call cord in each call cord jack. If the
call
placed
lamp is illuminated, press the
reset
button to reset it.
IR155B Patient Stations: If the call cord has been pulled, reset the station by pushing
the reset switch to the left.
SF155B Emergency or SF156B Code Stations: If the pull cord has been pulled, or the
switch had been pulled down, reset the station by pushing the switch up.
LI150B Duty Stations: Push the
tone
off
switch up.