3.8.6
Calibration Error Message (
Error SPAN
)
An error message indicates that the calibration process was not successful. After the error message
is displayed, the meter will revert to the 4.000 calibration menu. The error message might be
caused by inadvertently leaving the signal at the previous level or not maintaining the minimum
span. Press the
MENU
button to cancel the current calibration process if necessary.
3.8.7
Level Input Multipoint Linearization (
MULTIPT
)
This menu enables multipoint linearization for scaling and calibrating of the level display. Setting
MULTIPT
to
ENABLE
will alter the level display Scaling (See Scaling the Meter) and Level Calibration
(See Level Input Live Signal Calibration) menus to include a Number of Points (
NO PTS
) parameter
before entering Input 1.
32-point linearization can be used to linearize the display for non-linear signals.
3.8.8
Information (
INFO
)
The Information menu shows the software identification number and version number. To
determine the software version of a meter:
Go to the Information menu (
INFO
) and press
ENTER
button.
Continue pressing
ENTER
to scroll through the software release number and software version.
Following the information display, the meter will exit the Advanced Features menu and return to
run mode.