Operation Manual TCH Pro 202 & 203
•
As noise does not saturate the sensor
due to the particle detection filter, mono-
disperse aerosols may be combined to
execute a single-sample calibration.
•
Graph caption
: Median peak distribution
of 5 mono-sized NIST-traceable PSL
spheres is utilized as threshold for each
calibrated particle size. The image shows
peaks for 0.2µm, 0.3µm, 0.5µm, 1µm and
2.0µm. Right of the mean (shaded areas)
are included in the total count for each
specific particle size.
Page of
13
33
TCH Pro v1.2.1