1.3. Control Module
The control module is the interface between the RS485 sensor bus and data communication links. It
contains a display, a SD data card which logs data, a RS232 serial connector for external
communication, a bus connector for internal communication with the sensor modules and a cable
connector for the Sensiron Humidity and Temperature sensor (if fitted).
The Dust Profiler is supplied with a RS232 to USB adapter fitted between the Control Module and an
external USB socket fitted at the base of the enclosure. This enables the user to perform a wide range
of functions over the USB connection, without needing to open the enclosure door, such as data
logging and various system checks using the supplied Aeroqual Dust Profiler PC software.
Note 1: The driver for the RS232 to USB adapter is in the USB stick which comes with the
instrument.
Note 2: The power to sensor and alarm relays are also attached to the control module base
plate.
Display
Front
6 way Alarm Relay
SD Card Slot
DB9 RS232
4 way T/RH
2 way Bus relay
activation
RJ45 RS485
Bus
2.1mm 12VDC relay
switched Power output
Top
Bottom
Off/On switch with LED.
Depress to power on (blue
LED on)
Controller 12VDC
power input
Programming Port
RS232
Firmware (V6.2+)
Programming Dip
Switch (position is
important)
Firmware (V6.2+)