Note: These are EXAMPLE settings; instrument will have slight variations depending on
calibration and software version releases
Sensor Diagnostics View
Not applicable for the Dust Profiler. All readings will show blank.
Parameter
Value
Meaning
Version
5.6
Firmware version (software controlling instrument operation)
HTR
45
Reserved
H3
298.15
Ambient Temperature (K)
TEMA
1.4
Density 1 (PM1 measurement)
TIMA
2.00
Density 2 (PM2.5 measurement)
TEMR
2.00
Density 3 (PM10 and TSP measurement)
TIMR
60
Accumulation time
H0
0
Offset
H1
1
Volumetric flow rate (LPM)
H2
760
Ambient Pressure (Torr)
PWMH
35
Not used
PWML
60
Not used
Gain
1
Gain on reading (equivalent to gain on Control module)