![Teesing Aeroqual Dust Sentry Pro User Manual Download Page 21](http://html1.mh-extra.com/html/teesing/aeroqual-dust-sentry-pro/aeroqual-dust-sentry-pro_user-manual_1073612021.webp)
Select Setup
Email Recipients and enter the Email server and Email recipients
NOTE: the SMTP server will typically be one provided by your SIM card provider. The user
name and password are often not required.
3.3.3.
SMS Message
This function assumes the modem is fitted and correctly configured. It is activated by the
“Enable
SMS Alert
” tick box under Operations. Up to 5 SMS recipients can be connected at once.
Select Setup
Mobile Recipients and enter the relevant mobile phone numbers
Note: The country code needs to be included e.g. +44
3.4. Software Description Summary
The Dust Monitor Software is designed to be a simple interface for communicating with the Dust
Profiler monitor via a PC. It can be used to configure the instrument, initiate the zero calibration
routine, modify gain factors, poll data and display the data in either table or graphical format. The
software runs in Java VM1.5 (supplied with software) and the database is an open source HSQLDB
Java database. Data can be easily exported to a spread sheet for processing. The software also
incorporates a TCP/IP modem connection which can be used to connect to and operate a Dust
Profiler unit remotely.
3.4.1.
File
Search monitor
Searches for available Dust Profiler monitors
Connect to Monitor
Connects to Dust Profiler
Start Data Logging
Starts data logging from the Dust Profiler
Stop Data Logging
Stops data logging from the Dust Profiler