Operating Instructions

Pressure Gauges 

Model P2112

Model P2330

loadable up to 40 bar

Differential Pressure

Model P2680

loadable up to 400 bar

tecsis GmbH

Carl-Legien-Straße 40-44

63073 Offenbach . Germany

Tel.: +49 (0)69/5806 0 

Fax: +49 (0)69/5806 7788

E-Mail: [email protected]

www.tecsis.de

Notes according to Pressure Equipment Directive 

97/23/EC

 

The pressure gauges are "pressure accessories"  

 

       

in accordance with article 1, paragraph 2.1.4.

 

The volume of the pressure bearing housing of tecsis    

       pressure gauges is < 0.1 L.

 

The pressure gauges carry the CE marking for fluid    

 

group 1G in accordance with annex 2, table 1 when  

 

 

their permissible working pressure exceeds 200 bar.

 

Pressure gauges that do not carry the CE marking  

 

      are manufactured in accordance with article 3,  

 

      paragraph 3 "sound engineering practice".  

1 Safety instructions

 

The user must ensure that the appropriate  

 

 

pressure gauge with regard to scale range  

 

 

and performance and the appropriate  

 

 

wetted material (corrosion) for the specific 

  

measuring conditions of the respective appli-

cation is selected. In order to guarantee the accuracy and 

long-term stability specified, the corresponding load limits 

are to be observed. Specifications: see data sheet under 

www.tecsis.de

Only qualified persons authorised by the plant manager 

are permitted to mount, maintain and service the pressure 

gauges.

Dangerous pressure media such as Oxygen, Acetylene, 

flammable gases or liquids, toxic gases or liquids as well 

as for refrigerating plants or compressors requires atten-

tion above the standard regulations. Here the specific 

safety codes or regulations must be considered. 

After an external fire, pressure media can leak out, parti-

cularly at soft solder joints. All gauges must be checked 

and, if necessary, replaced before recommissioning the 

plant. 

Serious injuries and/or damage can occur should the 

appropriate regulations not be observed.  

2 Mechanical connection

According to the general technical regulations for pressure 

gauges, respectively (i.e. EN 837-2). When screw-fitting 

the gauges the force required for this must not be applied 

through the case or terminal box but just through the 

spanner flats (with suitable tool) provided for this purpose. 

   

Installation with

 

open-ended spanner

Correct sealing of pressure gauge connections with 

parallel thread 

 

shall be means of a suitable sealing 

ring, sealing washer or tecsis profile seals. The sealing of 

tapered threads (e.g. NPT threads) is made by providing the 

thread 

 

with additional sealing material like, for example, 

PTFE tape (EN 837-2).  

When a blow-out device is fitted to a pressure gauge, it 

must be protected against being blocked by debris and dirt.

With safety pattern gauges (see dial symbol        or s) you 

need to pay attention to the fact that the free space behind 

the blow-out back will be at least 15 mm.

2.1 Special requirements for the installation point

If the measuring point is not adequately stable a measuring 

instrument support such as a bracket or flange should be 

used for fastening (and possibly via a flexible capillary line). 

If the pressure gauge is exposed to vibration or pulsating 

pressure or both, then a liquid filled pressure gauge may 

provide considerably better performance and readabili-

ty. Instruments should be protected against coarse dirt 

and wide fluctuations in ambient temperature. EN 837-2 

„Selection and installation recommendations for pressure 

gauges“ should be complied with.

3 Admissible ambient and working temperatures

When mounting the pressure gauge it has to be ensured 

that, taking the influence of convection and heat radiation 

into consideration, no upper or lower deviation from the 

permissible ambient and medium temperatures can occur. 

The influence of temperature on the class accuracy is to be 

observed.

4 Storage

The pressure gauge should remain in its original  

packing until installation. The gauge should be  

protected from external damage during storage.

Storage temperature: -40 °C ... +70 °C.

Pressure gauges removed from service should be protected 

from dust and humidity.

5 Maintenance and servicing / Repairs

The instruments require no maintenance or servicing.  

Tests should be carried out on a regular basis to  

guarantee the measuring accuracy of the pressure gauge. 

The tests or recalibrations have to be carried out by quali-

fied persons with the appropriate  

equipment.

 

Remainder of the pressure medium  

 

 

contained in the pressure element may  

 

 

be hazardous or toxic. This should be  

 

 

considered when handling and storing  

 

 

the removed pressure gauge.

6 Applied standard

EN 837-1  Bourdon tube pressure gauges,  

 

dimensions, metrology, requirements and 

 

testing

EN 837-2  Selection and installation 

 

recommen  dations for pressure gauges

EN 837-3  Diaphragm and capsule pressure gauges,  

 

 

dimensions, metrology, requirements and  

 

testing

Technical alteration rights observed.

The torque depends on the seal used. With standard G-type 

pipe thread, gauge connection by means of a union nut or 

a LH-RH adjusting nut is recommended to simplify correct 

orientation of the gauge.

wrench surface

sealing of the 
thread

 

sealing area

 

2450547

BD/BE /BF 100 c         06/2011

ADPR1X100001