background image

                                                                                                                             

HUMIDITY TEMPERATURE METER 

 
 

 

4

TX

RX

GND

Press "MAX MIN" again, MAX, and MIN will blink together. This means that all these data is updated 
in the memory and the reading is the present temperature. 
One may press "MAX MIN" to circulate the display mode among these options.   
When the meter is under "MAX MIN" operation, “ 

°C 

“ “ 

°F 

” “ 

U

REL

 ”

 

button are disabled.(when you 

press “ 

°C 

“ “ 

°F 

”“ 

U

REL

 ”

 

button in "MAX MIN" mode, there will be two continuous beep) 

To exit the MAX/MIN mode, one may press and hold "MAX MIN" for two seconds. 

4.7 Relative Operation: 

When one press the“ 

U

REL

 ” button, the meter will memorize the present reading and the 

difference between the new reading and the memorized data will be shown on the display, Press the 
“ 

U

REL

 ” button again to exit the Relative operation. When the meter is under relative operation, 

“ 

°C/°F 

” button is disabled. ( when you press “ 

°C 

“ “ 

°F 

” button in relative mode , there will be two 

continuous beep )   

4.8 Auto Power Off: 

By default, when the meter is powered on, it is under auto power off mode. The meter will power 
itself off after 30 minutes if no key operation and no RS232 communication. 
Combination at power on can disable auto power off.   
One may press and hold “

HOLD

” button and then power on the meter and there will be two 

successive beeps to indicate that auto power off is disabled and the 

         

will not show up.

 

4.9 Low Battery Condition   

When the battery voltage is under proper operation requirement, the            symbol will show on the 
LCD and the battery need to be replaced with new one. 

4.10 Digital Output: 

The Digital Output is a 9600bps N 81 serial interface. 

 

 

 

1.Don’t touch or manipulate the sensor. 
2.Don’t expose the sensor to direct light , this causes a false reading. 
3.Don’t expose the sensor to static electricity. 

 

 

Appendix: Thermo couple probe specification   

Model Range 

Tolerances 

Description 

TP-K01 

Bead probe 

-50

 to 200

 

-58

 to 392

 

±

2.2

 or 

±

0.75

 

(

±

3.6

 or 

±

0.75

with Teflon tape insulation Maximum 
insulating temperature : 260

 

 

                                                         

 

TP-K01: 

probe for general condition measurements, especially for 

complex and hard to reach places.

 

WARNING!

 

TECPEL

Summary of Contents for DTM-322

Page 1: ...REL HOLD C F T1 K TYPE T2 POWER UP OPTIONS C F 0 TEMP 20 C HUMIDITY 328 200 4 F 60 C 2498 1370 RH 140 F 100 F C DTM 322 TECPEL ...

Page 2: ...311 01 JUL 2000 TECPEL ...

Page 3: ... Instructions 3 4 1 Power Up 3 4 2 Humidity and Temperature Measurement 3 4 3 Connection the Thermocouples T2 channel 3 4 4 Selecting the Temperature Scale 3 4 5 Data Hold Operation 3 4 6 MAX MIN Operation 4 4 7 Relative Operation 4 4 8 Auto Power Off 4 4 9 Low Battery Condition 4 4 10 Digital Output 4 VI Calibration Procedure 5 TECPEL ...

Page 4: ...r sensor and K type thermocouple This operations manual contains general product information and specification III Specifications Numerical Display 4 digital Liquid Crystal Display Measurement Range Humidity 0 100 RH Temperature T1 20 60 4 140 T2 200 1370 328 2498 Resolution Humidity 0 1 RH Temperature T1 0 1 0 1 T2 200 200 0 1 200 1370 1 200 200 0 1 else 1 Accuracy Humidity 2 5 RH at 25 Temperatu...

Page 5: ...on Meter 186 L 64 W 30 H mm 10 8 L 2 5 W 1 2 H inch Probe 190 L 15 D mm 7 5 L 0 6 D inch Accessories Instruction Manual 9V Battery Carrying Case Probe Holder K type bead thermocouple probe IV Symbol Definition and Button Location Centigrade and Fahrenheit indication RH Relative Humidity indication MAX The Maximum value is now being displayed MIN The Minimum value is now being displayed This indica...

Page 6: ... Fahrenheit F by pressing F button and vice versa to Celsius by pressing C button 4 5 Data Hold Operation The user may hold the present reading and keep it on the display by pressing the HOLD button When the held data is no longer needed one may release the data hold operation by pressing HOLD button again When the meter is under Data Hold operation the MAX MIN and C F UREL button are disabled whe...

Page 7: ...y default when the meter is powered on it is under auto power off mode The meter will power itself off after 30 minutes if no key operation and no RS232 communication Combination at power on can disable auto power off One may press and hold HOLD button and then power on the meter and there will be two successive beeps to indicate that auto power off is disabled and the will not show up 4 9 Low Bat...

Page 8: ...he user perform the humidity calibration the environment should be kept at the stable condition i e Constant temperature and constant humidity in the lab to in increase the accuracy 2 After the probe insert into the standard humidity cavity the operator should wait at least 20 minutes to let the condition in the cavity to stabilize 3 During the calibration if the user press POWER button at any tim...

Page 9: ... mode Remarks 1 After the probe is in the calibration cavity wait at least 20 minutes to stabilize the system 2 During the calibration the user can leave the process by pressing the POWER button at any time and the calibration data will be kept unchanged 3 During the calibration mode the user can restore the factory default value by press the HOLD and C buttons at the same time 4 Because it take s...

Page 10: ...www tecpel com tw www tecpel com TEL 886 2 2218 3111 FAX 886 2 2218 3222 Email sales tecpel com tecpelco gmail com TECPEL ...