ORDINARY, SCHEDULED AND EXTRA MAINTENANCE
ROUTINE MAINTENANCE:
Maintenance of the device consists simply of changing the filtering cartridges of the resin
container, which is to be carried out once a year.
CHANGING FILTERS AND RESINS:
1.
Disconnect the power connector on the front of the device. Close the water tap upstream of the
device.
2.
If the device is installed below the level of the autoclave, shut off the load on the back of the
autoclave itself to avoid emptying the tank of the autoclave and the subsequent flooding.
3.
Place the device on a flat surface and remove the lid. Use the spanner supplied (1NPUA0032) to
unscrew and also remove the PP sediment pre-filter cartridge, the GAC cartridge and the CTO
carbon-block filter cartridge (Fig. 1). Mop up the small amount of water that comes out with paper
or rags.
4.
Insert the cartridges in the containers and screw in the new filters (remove the transparent
protection (Fig. 2) and make sure you do not change their position. Screw them on tightly but do
not over-tighten (Fig. 3).
5.
Remove the safety clips from the fittings of the resin container (Fig. 4) and take out the connecting
pipes by pressing the retaining ring towards the body of the fitting and then pulling on the pipe.
6.
Remove the spent resin container and replace it with a new one, taking care to keep it the same
way round (Fig. 5).
7.
Insert the pipes in the fittings, press them home (Fig. 6) and insert the two safety clips (Fig. 7).
8.
Turn the two ball valves so that the levers are parallel to the pipes entering and leaving them (Fig.
11 & 12),
i.e. in the “open” position.
9.
Open the water tap upstream of the device and let the water run for at least 1 minute.
10.
Turn the two ball valves so that the levers are at right angles to the pipes (Fig. 13 & 14),
i.e. in the
“closed” position.
11.
Check that the final condition of the device resembles the one shown in Fig. 15 .
12.
Record label of the sediment filter the date of filters exchange of the device (Fig. 16 ).
13.
Check that no water is leaking out and put the device in its original position.
14.
Press and hold the button, load water manually on the control panel of the autoclave until a
beep.
ATTENTION:
Replacing the filters and resins is absolutely necessary after the signal from the
connected autoclave, or IN ANY CASE AFTER 12 MONTHS so as not to compromise the
functionality of the system and also to preserve the osmotic membrane.
Summary of Contents for OI002NPU
Page 2: ......
Page 21: ...19 ITALIANO FIGURE E SCHEMI Fig 1 Fig 2 Fig 3 Fig 4 1 2 3 ...
Page 22: ...20 ITALIANO Fig 5 Fig 6 Fig 7 Fig 8 Fig 9 Fig 10 ...
Page 23: ...21 ITALIANO Fig 11 Fig 12 Fig 13 Fig 14 Fig 15 Fig 16 ...
Page 45: ...19 ENGLISH ILLUSTRATIONS DIAGRAMS Fig 1 Fig 2 Fig 3 Fig 4 1 2 3 ...
Page 46: ...20 ENGLISH Fig 5 Fig 6 Fig 7 Fig 8 Fig 9 Fig 10 ...
Page 47: ...21 ENGLISH Fig 11 Fig 12 Fig 13 Fig 14 Fig 15 Fig 16 ...
Page 69: ...19 FRANÇAIS FRANÇAIS CHIFFRES HORAIRES Fig 1 Fig 2 Fig 3 Fig 4 1 2 3 ...
Page 70: ...20 FRANÇAIS Fig 5 Fig 6 Fig 7 Fig 8 Fig 9 Fig 10 ...
Page 71: ...21 FRANÇAIS FRANÇAIS Fig 11 Fig 12 Fig 13 Fig 14 Fig 15 Fig 16 ...
Page 73: ...23 FRANÇAIS FRANÇAIS Système électrique Fig 19 Système hydraulique Fig 20 ...
Page 93: ...19 DEUTSCH ABBILDUNGEN SCHEMEN Fig 1 Fig 2 Fig 3 Fig 4 1 2 3 ...
Page 94: ...20 DEUTSCH Fig 5 Fig 6 Fig 7 Fig 8 Fig 9 Fig 10 ...
Page 95: ...21 DEUTSCH Fig 11 Fig 12 Fig 13 Fig 14 Fig 15 Fig 16 ...
Page 97: ...23 DEUTSCH Elektrische Schema Abb 19 Hydraulische Schema Abb 20 ...
Page 117: ...19 ESPAÑOL FIGURAS ESQUEMAS Fig 1 Fig 2 Fig 3 Fig 4 1 2 3 ...
Page 118: ...20 ESPAÑOL Fig 5 Fig 6 Fig 7 Fig 8 Fig 9 Fig 10 ...
Page 119: ...21 ESPAÑOL Fig 11 Fig 12 Fig 13 Fig 14 Fig 15 Fig 16 ...
Page 121: ...23 ESPAÑOL Esquema eléctrico Fig 19 Esquema idraulico Fig 20 ...
Page 123: ......