- 47 -
7.01
Apply Power to the System
Assuming that all the entered Parameters are correct and all the equipment is installed properly,
the scale must now be calibrated. Without calibration, the scale will only approximate the RATE.
First, however, on scales with LVDT(s) (
L
inear
V
oltage
D
ifferential
T
ransformer), a procedure
needs to be performed to verify that the LVDTs are in alignment. Use one of the following two
sections, depending upon whether the scale has a SC200 or SC400 signal conditioner, to verify
(and re-align if necessary) the LVDT alignment.
Now turn on the conveyor, power up the WP20, make sure the belt is clean, and let the entire
system run for at least half an hour (longer in colder environments) so it attains temperature
stability. If the WP20 displays a "Scale Data Error", there’s a communications problem between
the scale (the signal conditioner) and the WP20. This problem must be corrected before
continuing. Consult Section 9.08, “scale data error”, for troubleshooting tips.
7.02 LVDT Alignment for Scales with a SC400 Signal Conditioner
(WY10, WY10HD, WY20 scales Only)
The WY10 and WY10HD have two LVDTs, while the WY20 has four LVDTs. Alignment of the
LVDTs is critical for scale accuracy. The LVDTs were aligned at the factory, but it must be
verified that they are still in alignment after shipping and installation. Turn the belt off, leave the
MS20 on and remove both side covers from the scale. On the SC400 PCB, DIP switches SW3
and SW4 are used in the alignment procedure (see the figure below). First verify that SW3 and
SW4 are set to the positions shown on the first or second lines in the following chart depending
on the particular scale at hand. Next, to verify proper alignment, connect a DC voltmeter across
the test points (again see the figure). If the voltage is 2.50 +/- 0.02 VDC, the LVDTs are
sufficiently aligned and move on to the next section, however, if the voltage is outside the stated
range, continue as follows.
CHAPTER 7 – CALIBRATION
Summary of Contents for Tecweigh 20
Page 2: ... 1 ...
Page 10: ... 9 2 09 I O Quick Reference Guide ...
Page 20: ... 19 ...
Page 24: ... 23 4 03 WY15 and WY25 Scale Wiring load cells ...
Page 25: ... 24 4 04 HY15 and HY25 Field Wiring ...
Page 26: ... 25 4 05 WY10 WY10HD and WY20 Scale Wiring LVDTs ...
Page 27: ... 26 4 06 Chart Recorder Data Logger using the 4 20 ma output ...
Page 28: ... 27 4 07 Chart Recorder Data Logger using the HFR output ...
Page 29: ... 28 4 08 Remote Rate Total Display 4 09 Remote Tons Counter ...
Page 30: ... 29 4 10 Radio Transceivers Local Scale Remote Processor ...
Page 31: ... 30 4 11 Radio Transceivers Local Processor Remote PLC PC or Remote Display ...
Page 32: ... 31 4 12 Radio Transceivers Local DC Generator Remote Processor ...
Page 33: ... 32 4 13 Radio Transceivers Local DC Generator and Processor ...
Page 34: ... 33 4 14 Multiple Processors to Tecweigh Multi Scale Display MSD ...
Page 61: ... 60 SC300 PCB WY15 and WY25 SC400 PCB WY10 and WY20 ...
Page 71: ... 70 10 01 WY10 Dimensions CHAPTER 10 DIMENSIONS ...
Page 72: ... 71 10 02 WY10 HD Heavy Duty Dimensions ...
Page 73: ... 72 10 03 WY15 Dimensions ...
Page 74: ... 73 10 04 WY20 Dimensions ...
Page 75: ... 74 10 05 WY25 Dimensions ...