65
DEVELOPMENT, MANUFACTURING AND SUPPLY OF INSTRUMENTATION
Appendix 4
Transmitter AG-105 Technical specifications
The frequencies of the continuous “Co” or pulse “Pu” signal, Hz ± 0,1% - “kHz”
Load.
“clips”
.or.
“clamp”
512
-.“0.5”./.
1024
.-.“1.0”./.
8192
.-.“8.2”./.
32768
.-.“33”
“Antenna”
.modes
8192
-.“8.2”./.
32768
.-.“33”
Operating modes
“Antenna”
.mode
Internal.transmitting.inductor.“
In
”.
“Modulation”
.modes.(special.
form.signal)
Pulse.“
Pu
”.(short-term.transmissions.of.the.sine.signals).
Transmission.duration.0,12sec.
Transmissions.repetition.frequency.1Hz.
Dual-frequency.“
2F
”.(simultaneous.frequency.generation.1024Hz,.8192Hz).
Amplitudes.ratio.4/1.(respectively).
Output parameters under power supply voltage 12…15V
Output current, А
Restricted.by.the.program.
under
manual increment,
≥
5
–.at.frequency.512Hz.“0.5”./.1024Hz.“1.0”./.8192Hz.“8.2”./.“2F”
3
-.at.frequency.32768Hz.“33”
Set.by.the.program.for.
automatic adjustment
.with.
the.external.load.of.“clips”.or.
“clamp”,.
≥
0,2
.–.at.frequency.512Hz.“0.5”./.1024Hz.“1.0”./.“2F”
0,1
.–.at.frequency.8192Hz.“8.2”./.32768Hz.“33”
Maximum output voltage, V
Depending.on.“modulation”,.
≥
32
.–.in.the.dual-frequency.modulation.mode.“2F”
40
.–.in.other.modes
Maximum output power, W
Restricted.by.the.program,.
≥
20 -
.In.the.continuous.“Co”.and.pulse.“Pu”.modes.at.frequencies.512Hz.“0.5”.
/.1024Hz.“1.0”./.8192Hz.“8.2”.at.load.resistance.up.to.80
Ω
In.dual-frequency.mode.“2F”.at.load.resistance.up.to.50
Ω
..
6
.-.At.frequency.32768Hz.“33”.at.load.resistance.up.to.260
Ω
.
Power supply sources
Operating.range.of.the.supply.
voltages
Minimally.acceptable.voltage.for.oscillation.start-up:.
7V
.“Bt”,.<9.V.“Ас”.
Maximum.allowed.voltage.for.operation.–.
15V.
Automatic.shutdown.voltage.in.the.“oscillation”.mode:.<.
4,2V
“Bt”,.<7,9.V.“Ас”
Battery
.set
8.“alkaline”.cells.1,5V.“type.С”
Recommended.–.“Duracell.®.ULTRA”.
External
.power.supply.sources.
(optional)
Battery
“12V”.(e.g..automotive)
Output.voltage.11…14V,.maximum.current.not.less.than.4А.
Mains power supply adapter
АG114М.02.020.(extra.accessory.based.on.
GS60A15-P1J.“MEAN.WELL”)
Output.voltage.15V,.power.60W.
Operating time (“life cycle”)
While.in.operation.of.the.battery.set.“type.Сх8”.is.defined.by.the.quality..
(capacity.and.“load.capability”).of.the.applied.alkaline.cells.and.can.be.equal..
to
from 4 to 6 hours in the “Co” and “2F” modes or from 20 to 30 hours
in the “Pu” mode
.at.initial.output.voltage.7W.in.“continuous”.modes.“Co”./..
“2F”.or.at.initial.output.voltage.15W.in.pulse.modulation.mode.“Pu”.
Under.the.external.power.supply.is.fully.conditioned.by.this.external.energy..
source.characteristics.therefore
under mains supply operating time is
unlimited.
Summary of Contents for AG-309.15G