![TECHNO-GAZ Master Flux Plus Instructions For Use Manual Download Page 48](http://html.mh-extra.com/html/techno-gaz/master-flux-plus/master-flux-plus_instructions-for-use-manual_1070059048.webp)
ENGLISH
20
END OF TREATMENT
The treatment will end when the orthodontist has finished the procedure on the patient.
To end treatment:
-
Shut off the NITROUS OXIDE supply by turning (counter-clockwise) the %N
2
O knob
(C – figure A10) to the end stop (0%).
-
Administer (as at the beginning of treatment) a flow of 3 Lt/min of oxygen for 2
minutes
-
Shut off the OXYGEN supply by turning (counter-clockwise) the FLOW knob (A –
figure A10) to the end stop.
- Loosen the loop that secures the tube and take the mask off of the patient.
Before signing the patient out keep him/her under observation (waiting room) for
~5÷10 minutes.
The doctor may deem it necessary to increase the observation period.
CLOSING THE CYLINDERS
Once the doctor has finished using the device it is IMPORTANT to close the cylinders.
Close the valve on the Nitrous oxide cylinder (N2O), (turning it clockwise). Close the
valve on the Oxygen cylinder (O2), (turning it clockwise).
Summary of Contents for Master Flux Plus
Page 11: ...ITALIANO 9 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figura 3 ...
Page 12: ...ITALIANO 10 2 F 10 11 12 13 14 15 Figura 4 ...
Page 36: ...ENGLISH 8 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figure 3 ...
Page 37: ...ENGLISH 9 2 F 10 11 12 13 14 15 Figure 4 ...
Page 62: ...FRANÇAIS 8 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figure 3 ...
Page 63: ...FRANÇAIS 9 2 F 10 11 12 13 14 15 Figure 4 ...
Page 86: ...DEUTSCH 6 1151 mm 516 mm 428 mm Bild 1 900 mm 1305 mm 230 mm 500 mm max 400 mm 200 mm Bild2 ...
Page 89: ...DEUTSCH 9 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Bild 3 ...
Page 90: ...DEUTSCH 10 2 F 10 11 12 13 14 15 Bild 4 ...
Page 116: ...ESPAÑOL 9 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figura 3 ...
Page 117: ...ESPAÑOL 10 2 F 10 11 12 13 14 15 Figura 4 ...
Page 134: ......
Page 135: ......