ENGLISH
15
6.4
Evacuation of exhaled gases:
To safeguard the work environment and the staff, the exhaled gases must be conveyed
outside of the work environment.
Two systems can be used to discharge the gases externally:
- Direct connection to the outside of the office
- Connection to a surgical aspirator.
It is important to evaluate:
- Whether the aspirator is suitable for the evacuation of the gases (
N
2
O and O
2
)
- The aspirator exhaust must be outdoors and ventilated.
Make sure it is compliant with regulations in force in the country of use.
Direct connection to the outside of the office
UNIT-MOUNTED MASTER FLUX
Screw the nozzles
(Figure 3 Pos 8/9)
onto the unit.
Remove the supplied valve
(Figure 5 Pos 17)
.
Connect the mask circuit gas exhaust tube to the nozzle
( Figure 3 Pos 9)
.
Connect the gas evacuation spiral tube
(Figure 9)
to the nozzle
(Figure 3 Pos 8)
on the unit.
Connect the end of the spiral tube to the outside and make sure there is a final filter.
WALL-MOUNTED MASTER FLUX
Connect the gas exhaust tube to the mask circuit to the nozzle
( Figure 9 Pos 36)
.
Connect the end of the spiral tube to the outside and make sure there is a final filter
(Figura 9 Pos 34)
.
Summary of Contents for Master Flux Plus
Page 11: ...ITALIANO 9 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figura 3 ...
Page 12: ...ITALIANO 10 2 F 10 11 12 13 14 15 Figura 4 ...
Page 36: ...ENGLISH 8 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figure 3 ...
Page 37: ...ENGLISH 9 2 F 10 11 12 13 14 15 Figure 4 ...
Page 62: ...FRANÇAIS 8 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figure 3 ...
Page 63: ...FRANÇAIS 9 2 F 10 11 12 13 14 15 Figure 4 ...
Page 86: ...DEUTSCH 6 1151 mm 516 mm 428 mm Bild 1 900 mm 1305 mm 230 mm 500 mm max 400 mm 200 mm Bild2 ...
Page 89: ...DEUTSCH 9 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Bild 3 ...
Page 90: ...DEUTSCH 10 2 F 10 11 12 13 14 15 Bild 4 ...
Page 116: ...ESPAÑOL 9 1 2 3 4 5 6 7 9 8 F 7 6 5 4 Figura 3 ...
Page 117: ...ESPAÑOL 10 2 F 10 11 12 13 14 15 Figura 4 ...
Page 134: ......
Page 135: ......