Diagnostics and troubleshooting
Proline Promass 500 Modbus RS485
28
Error
Possible causes
Solution
Unsuitable view settings.
Change the font size/display ratio
of the Web browser.
No or incomplete display of
contents in the Web browser
• JavaScript not enabled
• JavaScript cannot be enabled
1. Enable JavaScript.
2. Enter http://XXX.XXX.X.XXX/
basic.html as the IP address.
6.2
Diagnostic information in the Web browser
6.2.1
Diagnostic options
Any faults detected by the measuring device are displayed in the Web browser on the
home page once the user has logged on.
2
3
1
A0031056
1
Status area with status signal
2
Diagnostic information
3
Remedy information with Service ID
In addition, diagnostic events which have occurred can be shown in the
Diagnostics
menu:
• Via parameter
• Via submenu
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
Symbol
Meaning
Failure
A device error has occurred. The measured value is no longer valid.
Function check
The device is in service mode (e.g. during a simulation).