OPERATION

To set the temperature controller to the required operating temperature, switch on

the electrical supply which will illuminate the power and cut-out indicators.  Then set the
required temperature by using the up and down arrows.  Depress the reset switch to initiate the
system and extinguish the cut-out indicator.

Your unit is fitted with a 

Eurotherm 2208L controller

 and the parameters are

factory set and should not be changed.

As the fluidised bath heats up the fluidisation will become more vigourous, due to the

expansion of the fluidising air.  Units fitted with the automatic fluidising air control system will
automatically reduce the air input so as to keep a constant level of fluidisation.  Manually
operated units must the fluidising air flow, reduced unit the level of fluidisation returns to its
origional state.  The following table indicates the flow rates that are factory set on an auto-air
unit

Ambient to 200°C

300 l/min

200°C   to 400°C

180 l/min

400°C   to 600°C

130 l/min

The fluidised bath should be operated with the bath lid fitted to reduce heat loss

during the heat up period, and to increase the efficiency of the fume extraction system.  As the
temperature of the fluidised bath approaches the set point, the heater indicator, situated in the
temperature controller fascia, will cycle on and off, until the actual temperature reaches the set
point temperature.

It may be desirable, depending on the process being carried out in the fluidised bath,

to check the actual bath temperature with an independent temperature sensor.  Any offset in the
temperature may be corrected by adjustment of the temperature controller (an instruction
manual for the temperature controller is included with this manual).  Further temperature
checks may be made at various points within the fluidised bath to ensure good temperature
uniformity is being achieved throughout the bed.  Slight adjustment of the fluidising air flow rates
will reduce any excessive temperature gradients.

CAUTION: Care should be taken when handling hot parts which have been removed

from a fluidised bath. We recommend that protective clothing (gloves, aprons, and visors, etc)
be worn at all times when working with fluidised baths and that the installation and maintenance
procedures outlined in this booklet be followed explicate.

Summary of Contents for IFB-131

Page 1: ...IFB 131 OPERATOR SMANUAL Issue 8 Date of issue 3rd May 2005 Techne is a trademark Techne 2004...

Page 2: ......

Page 3: ...7 OPERATOR SAFETY 8 Warning Note 8 General Description 9 Technical Specification 11 INSTALLATION 12 Fume ignition 14 Fume Cleaning 15 OPERATION 16 Operating Procedures for Plastic Residue 18 MAINTENAN...

Page 4: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 2...

Page 5: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 3 NOTES...

Page 6: ...4 Place the unit on a suitable bench or flat workspace or in a fume cupboard if required ensuring that the air inlet vents on the underside are free from obstruction 5 Symbols on or near the power sw...

Page 7: ...ppareil sur un plan de travail ou surface plane ou le cas ch ant dans une hotte d aspiration en s assurant que les trous d a ration situ s sous l appareil ne sont pas obstru s 5 Les symboles situ s su...

Page 8: ...eitung Bitte lesen Sie diese Bedienungsanleitung komplett bevor Sie dieses Ger t benutzen Warnung HOHETEMPERATURENSINDGEF HRLICH sie k nnen dem Bediener ernsthafte Verletzungen zuf gen und brennbare M...

Page 9: ...or del equipo 3 Conecte el cable a la toma de tensi n en la parte posterior del equipo 4 Sit e el aparato en un lugar apropiado tal como una superficie de trabajo plana o si fuera necesario incluso en...

Page 10: ...avoid violent fluidisation which occurs when damp alumina is heated above 100 C operate the bath for a period of approximately 8 hours at 90 C prior to operation at elevated temperatures NEVER ADD CO...

Page 11: ...sing atmosphere is required nitrogen can be utilised and if a reducing atmosphere is required cracked gas can be employed with a silicon carbide bed The unique characteristics of gas fluidised particl...

Page 12: ...is reduced and held at constant value until the temperature of the bath exceeds 400 C at which point a further adjustment is made A non standard version of this industrial fluidised bath is available...

Page 13: ...170 1030kPa 25 150psi Air consumption maximum 300 l min at 450 C 150 l min Nominal Heater Power 18KW Heatertype 9 ceramic knuckle heating element Control system 3 term PID temperature controller ther...

Page 14: ...ised the distance between the top of the aluminium oxide and the top of the bath is 390mm Should the fluidised bath be stored for long periods of time under damp or humid conditions moisture maybe abs...

Page 15: ...this manual ensuring that the fluidising air flow rate is adjusted in accordance with the above table b Switch on the fume extraction system c Immerse a typical component that is to be treated into th...

Page 16: ...the temperature of the bath is at minimum level to which satisfactory results can be achieved Increased working temperatures result in increased fume concentration By way of example an extraction rat...

Page 17: ...tain inorganic pigments or fillers such as titanium oxide of a particle size less than 0 005mm diameter which will not be retained by the cyclone it is recommended that a filtration system be fitted a...

Page 18: ...erated with the bath lid fitted to reduce heat loss during the heat up period and to increase the efficiency of the fume extraction system As the temperature of the fluidised bath approaches the set p...

Page 19: ...acted and treated by suitable fume handling equipment During the last third of the process time the polymer is reduced to carbon which either burns away or remains loosely bound to the object being cl...

Page 20: ...danger of the polymer melting and sinking to the bottom of the bath where it may block the porous plate With some large objects it may be desirable to preheat the bath to well above the operating temp...

Page 21: ...er elements as necessary With exceptionally dirty or wet air supplies this frequency may have to be increased Free water or water vapour in the air supply is a notorious source for the production of h...

Page 22: ...of corrosive alkalis such as caustic soda sodium hydroxide and quick lime calcium oxide should be avoided C MONTHLYINTERVALS 1 Empty the bath and inspect it for signs of corrosion especially pitted su...

Page 23: ...ater elements remove the inner container from the outer wrapper then remove the three layers of thermal insulation that surround the fluidised bath The heater elements are of the ceramic knuckle type...

Page 24: ...ated residue in the bath to be decomposed However if the fault continues the bath should be emptied and the porous plate examined The fluidised bath must only be emptied when the bath is cold To empty...

Page 25: ...007578 6007578 6007578 6007578 6007578 6006090 6006090 6005098 6005098 6005098 6005680 6005680 N A N A 6005680 6007767 6007669 6007669 6007668 6007670 6008457 6008457 6008457 6008457 6008457 6008321 6...

Page 26: ...5243 Extraction Fan 380 415V 50Hz F5148 Extraction Fan 380 415V 50Hz F6110 Reducing Spigot 150 100mm AB 100 Afterburner SR 100 Scrubber SRD 100 Caustic Doser CN 500 Cyclone Note that when the AB 100 a...

Page 27: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 25...

Page 28: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 26...

Page 29: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 27...

Page 30: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 28...

Page 31: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 29...

Page 32: ...I F B 1 3 1 O P E R A T O R S M A N U A L PAGE 30 Notes...

Page 33: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: