6.1 Minimum
distances
It is important to take in account some distances that must be kept in order to remove the
float due to a change in density or for maintenance purposes.
Distance LMS between the lowest side of the level gauge and the floor must be kept longer
or equal to LD distance.
On the other hand, the lower dimension LD, LP or LPV of series LT level gauges is variable
depending on the working liquid density. The lower the density, the longer the dimension.
Model
Liquid
density (kg/l)
Lower dimension
Without
drain (LD)
With drain
(LP)
With drain +
valves (LPV)
Without
vent (LS)
With vent
(LV)
With vent +
valve (LVV)
LT … LTL /
AISI 316L
0,55 ... 0,59
0,60 ... 0,91
≥ 0,92
430
340
260
445
355
275
590
500
420
130 155 300
LT … LTL /
PVC
0,60 ... 0,79
0,80 ... 0,89
≥ 0,90
400
310
240
525
435
365
150
140
265
LT ... LTL /
PP
≥ 0,70
240
365
150
165
290
LT … LTL /
PVDF
0,80 ... 0,89
0,90 ...0,99
1,00 … 1,19
≥ 1,20
415
340
290
240
540
465
415
365
150
165
290
Upper dimension