5 - Operation
Putting into Operation
Tecan Vacuum Separator (Te-VacS™) Operating Manual, 391236, en, V4.0
31 / 54
5.2.5 Connecting an Additional Waste Bottle
To connect an additional waste bottle (e.g., for foamy solutions) proceed as
follows:
1. Connect the standard waste bottle to the Te‑VacS tower using two tubes of the
same length (A, B).
A
B
C
E
A
D
E
B
C
D
Illustration 17:
Waste bottle connections
2. Connect the two waste bottles using a tube that is at least 5 m long (C).
3. Connect the additional waste bottle to the Te‑VacS tower using a tube of
suitable length (D).
4. Connect the liquid level sensor cable (E) to the vacuum unit.
5. Carefully tighten the tubing’s fixing nuts using a wrench.
NOTICE! It is recommended to tie the tubes together using cable ties.