background image

5

 

5.2 

Manually Controlling the Fan

Gradually increase the fan speed by slowly turning the fan controller clockwise. As 

the fan speed increases, the building depressurization displayed on Channel A should 

also increase. Continue to increase the fan speed until the building depressurization 

shown on Channel A is between –45 and –55 Pa.

 

 

 

 

 

 

 

 

 

 

6.  Channel B will display the one-point 50 Pa leakage estimate. Record this number.    

 

If the leakage estimate is fluctuating more than desired, try changing the time  

 

 

average setting on the gauge by touching the Time Average menu and choosing    

 

the 5 or 10 second or long-term averaging period. 

 

 

 

 

7.  Turn off the blower door fan. If you are using cruise control, this is done by touching the red stop icon in the Gauge app. 

8.  “-----” or “LOW”  appearing on Channel B

•  Whenever “-----” or “LOW”  appears on Channel B in the Flow @ 50 mode, the gauge can not calculate a reliable leakage 

estimate. The messages “-----”  and “LOW” appear on Channel B under the following conditions:

 

»

“-----” is continuously displayed when the building test pressure from Channel A is below a minimum value of 10 Pa. 

Estimating building leakage results when the test pressure is below this value may result in large errors. If possible, 

install a larger flow ring or remove the flow rings to generate more fan flow. Be sure the fan is off when changing 

flow rings.

 

»

Channel B reads “LOW” is continuously or LOW alternates with a flow reading when the air flow reading through 

the device is unreliable (i.e. you are trying to measure a flow outside of the calibrated range of the test device in 

its current configuration). If possible, you should change the test device configuration to match the flow rate being 

measured (e.g. install a flow ring or a smaller flow ring). Be sure the fan is off when changing flow rings.  

USING THE DG-1000 WITH THE MINNEAPOLIS BLOWER DOOR

Reviews: