NOTE
:
If the InnoMake Clip behaves in an incomprehensible manner or there are persistent,
unexpected malfunctions, the dealer must be contacted.
Malfunctions
Possible reasons
Necessary actions
An obstacle is
detected incorrectly
Foreign object on the sensor
Removal of the foreign object
Surrounding sources of interference
in the same frequency range of the
ultrasonic sensors, such as a
clanking bunch of keys, other
ultrasonic sensors
Restart and wait until it is ready
for operation
An obstacle is
detected too late
The measurement range was
changed just before the obstacle
was detected
Stop and stand still while making
changes to settings.
Complete loss of
function and
deactivation of the
InnoMake Clip
Device is overheated
Allow the InnoMake Clip to cool
down before touching it again;
avoid skin contact.
Battery is empty
Charge battery
Electrical or (electro-) magnetic
interference sources cause a failure
of the system
Restart the InnoMake Clip, the
settings and modes of operation
are retained
No warning tone
before automatic
deactivation
Software error
Contact dealer
No LED- feedback
while charging the
battery
Device may have overheated;
charging was automatically
terminated
Allow the InnoMake Clip to cool
down before touching it again;
avoid skin contact; charge when
cooled down.
USB interface defect
Contact dealer
Charger is not connected to power
Connect charger to power
Ready-to-operate
state is not
established
Software error
Restart the device, if necessary
charge the battery first