TDIMC1_4_User Manual_IMC-1.1.0_EN
17
NOTE
:
Abrupt changes in the environment may result in functional limitations or unreliable
feedback.
5.1.
Notes and measures in case of (mal)function of the product
Tipps, how to react to (mal-)functions of the InnoMake Clip.
Malfunctions
Possible reasons
Necessary actions
Dirt on the sensors
Cleaning the InnoMake Clip see
Obstacle is out of the range of 0,3
and 4,0 meters
Approach to or distance from an
obstacle in the walking direction
to bring it into the measuring
range and obtain appropriate
feedback
An obstacle is not
detected
The material or shape of an object
can cause unfavourable ultrasonic
reflections, so that it is not
detected as an obstacle.
Pay attention to the proper
execution of the primary aid
during the application of
InnoMake Clip.
Ultrasonic sensors are covered by
clothing
Do not wear clothing that
interferes with vibration feedback
or obstacle detection
Obstacle lies centrally between the
ultrasonic cones
Pay attention to the proper
execution of the primary aid
while using the InnoMake Clip
An obstacle is
detected incorrectly
At a larger measuring range, objects
located to the side of the selected
walking direction may be detected,
as the measuring range of the
ultrasonic sensors becomes wider
in a conical shape in the distance.
Limitation of the measuring range
of the InnoMake Clip to 1.5
meters in order to reduce the
width of the ultrasonic cones
Loading-LEDs (red
and blue) light up
simultaneously
Hardware error
Contact dealer