56
10-2-2. D/A Zero
Set the indicator value that you want to output for D/A Zero
(0V voltage).
10-2-3. D/A Full Scale
With the “10-2-2. D/A Zero” value as the reference, set the span
for the indicator values output to D/A.
When the “10-2-2. D/A Zero” and “10-2-3. D/A Full Scale” setting
values are added, the “10-2-4. D/A Max. Voltage” setting value
voltage is output.
The following is an example of when “10-2-4. D/A Max. Voltage”
is set to 1 V.
Setting example 1
D/A Zero
000.00
D/A Full Scale
100.00
Indicator value
D/A output
100.00
1 V
0.00
0 V
−100.00
−1 V
Setting example 2
D/A Zero
020.00
D/A Full Scale
100.00
Indicator value
D/A output
120.00
1 V
20.00
0 V
−80.00
−1 V
Setting example 3
D/A Zero
020.00
D/A Full Scale
−100.00
Indicator value
D/A output
120.00
−1 V
20.00
0 V
−80.00
1 V
Setting example 4
D/A Zero
−010.00
D/A Full Scale
020.00
Indicator value
D/A output
10.00
1 V
−10.00
0 V
−30.00
−1 V
ATTENTION
If calibration is conducted, when the rated capacity is con-
firmed, it will be set as the D/A Full Scale value.
10. System settings 1