![TE Technology TC-24-25 Operation Manual Download Page 13](http://html1.mh-extra.com/html/te-technology/tc-24-25/tc-24-25_operation-manual_3912188013.webp)
Expansion Connector Wiring Diagram
Note: The 25 mA current source is a true current source. Maximum compliance voltage is approximately
V+, the input voltage to the temperature controller. No external current limit resistors are needed for the
LED. The LED must be capable of being driven with a 25 mA continuous current.
8
7
6
5
4
3
2
1
Expansion Connector
Molex 22-23-2081
25 mA Current Source
Alarm Cancel
Input Common
High Alarm
Low Alarm
No Alarm
JP2
8
7
6
5
4
3
2
1
Customer- supplied
Connector
Note: Pins 6 and 7 not functionally defined.
Please do not connect.
8
7
6
5
4
3
2
1
Expansion Connector
Molex 22-23-2081
25 mA Current Source
Alarm Cancel
Input Common
High Alarm
Low Alarm
No Alarm
JP2
8
7
6
5
4
3
2
1
Customer- supplied
Connector
Note: Pins 6 and 7 not functionally defined.
Please do not connect.
13