background image

Created with the Personal Edition of HelpNDoc: 

Full-featured EPub generator

Other

Other is used by labs other than Euston, Gibb, or Metz. The file naming format follows the same method as
Euston, but files are saved in C:\Other\.

Created with the Personal Edition of HelpNDoc: 

Full-featured EBook editor

Continuous vs Timed Acquire

By toggling the Continuous checkbox, the user can determine whether the recording will time out
automatically after the number of seconds specified by the Recording Time field, or will run until the the Stop
Recording button is pressed. 
When Continuous is not toggled, the user can set the length of the recording by typing in the number of
seconds in the Recording Time field. Recording is not permitted when an invalid time is entered, as marked
by a red field background. A valid time must be a natural number (a whole number greater than 0).

Created with the Personal Edition of HelpNDoc: 

News and information about help authoring tools and

software

Storage Fidelity

The user has three bit depth options: 32-bit int, 24-bit float, or 16-bit float. All three are compatible with
Raven Pro. Higher bit depth settings can record smaller differences between sounds, but require more
memory.
Audio files can be downsampled after recording, but never upsampled, so it is recommended that users
choose 32-bit unless they are certain that a lower bit depth fulfills their requirements.

Created with the Personal Edition of HelpNDoc: 

Create iPhone web-based documentation

Start/Stop

Pressing the green "Start Recording" begins the acquisition of audio, which will display in the 

Waveform and

Spectrogram

 axes and write "Recording..." in the 

Status Output

. The button will change to the red "Stop

Recording" button which can be pressed at any time to halt acquisition, even when the timer has not fully
elapsed in the case of non-continuous recording. 

Created with the Personal Edition of HelpNDoc: 

Qt Help documentation made easy

Waveform & Spectrogram

Summary of Contents for RX6

Page 1: ...Ultrasonic Recording Manual...

Page 2: ...Stop 9 Waveform Spectrogram 9 Status Output 10 New Test Subject 11 Errors 12 Invalid field prevents record 12 Transfer rate is too slow 12 RX6 not connected 12 Microphone not connected 12 Program or...

Page 3: ...ncounter issues that are not addressed in this manual please contact Erica Nordin at erica nordin uleth ca Created with the Personal Edition of HelpNDoc Easily create PDF Help documents Starting the R...

Page 4: ...e desktop If the shortcut is absent the program can be found at C Ultrasonic Recording Program Ultrasonic_Recording_Interface Opening the program will display the main GUI outlined in the next section...

Page 5: ...recordings The microphone should be turned off between program uses The RX6 can be left on if no recordings are planned within the next week Created with the Personal Edition of HelpNDoc Write eBooks...

Page 6: ...Selecting a lab will bring the user to a new window with entry fields for data relevant to the file name The user can switch between the fields by pressing Tab and finalize their data by pressing Ent...

Page 7: ...with the Personal Edition of HelpNDoc Generate Kindle eBooks with ease Gibb When the Gibb automatic naming format is chosen the entry fields are Rat ID or Litter Number Postnatal Day and Experiment U...

Page 8: ...er 3rd 2017 at 4 30 pm for Rat 63 on Day 8 belonging to Cohort 4 the file will be named R63D8C4_2017 09 03_T16 30 wav Upon data entry another window will open asking whether the rat belongs to an expe...

Page 9: ...with the Personal Edition of HelpNDoc News and information about help authoring tools and software Storage Fidelity The user has three bit depth options 32 bit int 24 bit float or 16 bit float All th...

Page 10: ...a lower dB level The spectrogram displays vocalizations up to 100 kHz in frequency Due to limitations in the processor frequencies above 100 kHz cannot be represented in the recording Created with th...

Page 11: ...e program to enable a new recording After one recording is complete the program will not allow the user to press Start Recording again until it is initialized for the new recording task The current di...

Page 12: ...off the RX6 and microphone restart the PC and begin the start up sequence from scratch Created with the Personal Edition of HelpNDoc Free help authoring environment Microphone not connected The progr...

Page 13: ...t of the card and the critical failure is immediately rectified by a reboot the developers have decided to ignore this issue for now If this problem expands the PCI card will likely have to be replace...

Page 14: ...rd which the fiber optic cables connect to at the back of the PC Note that the currently installed card causes some issues upon booting the computer and may have to be replaced in the future See http...

Page 15: ...le and the text files mccExcludedFiles readme and requiredMCRProducts 2 The folder RecoverAudio containing the BinaryToWav executable and the text files mccExcludedFiles readme and requiredMCRProducts...