![TDT Medusa4Z User Manual Download Page 80](http://html2.mh-extra.com/html/tdt/medusa4z/medusa4z_user-manual_3900631080.webp)
Software Updates
In order to use the Medusa4Z you must have the TDT Drivers, BioSigRZ, and all other TDT
software at least at
v95
. Updates are available on our
fi
nd anything is
out of date.
Hardware Updates
In order to use the Medusa4Z your RZ6
microcode
must match your BioSigRZ version. The
update instructions are also on our Downloads page next to the TDT Drivers/RPvdsEx
.
Software Con
fi
guration Changes
For mouse models with the Medusa4Z, you will need to make a few changes to the stock
"Root_ToneABR.acf"
fi
le:
From the menu bar click Setup
→
Acquisition. In the Acquisition Setup screen change
the Source radio button from Bioamp to
Medusa4Z
and set the drop down menu to
12kHz
. See below indicated by the red arrows.
Immediately below the Source setup you will see a "Chan-1" button. Press this and
change the gain from
20
to
1
. See below indicated by the green arrows.
1.
a.
b.
Upgrading from RA4PA to Medusa4Z | 80