20 |
P a g e
G e n e r a l D a t a H e a l t h c a r e , I n c .
LED Indicators
•
TRIM LED – Indicates the microtome status at trimming and the
trimming thickness setting can be adjusted using the hand pad
interface.
•
SECT LED – Indicates the microtome status at sectioning and the
sectioning thickness setting can be adjusted using the hand pad
interface.
•
RET On LED – Indicates the microtome status at retraction and
retraction thickness setting can be adjusted using the hand pad
interface.
•
RUN LED - Indicates the microtome is automatically sectioning in
either Continuous or Single mode. The only adjustment that can be
made when this light is illuminated is the speed adjustment of the
sectioning.
Buttons for adjusting Section, Trim and Retraction values
•
Press the + (plus) or – (minus) buttons to set the section, trim, or
retraction value.
Adjusting range
: 0.5 to 600µm
0.5 to 2µm
,
0.5µm increments
2 to 10µm
,
1µm increment
10 to 20µm
,
2µm increments
20 to 100µm
,
5µm increments
100 to 600µm
,
50µm increments
Retraction range
:
From 5µm to 100µm, in 5µm increments.
Buttons for adjusting automatic sectioning speed.
•
The two buttons can be used while the unit is automatically
sectioning. The sectioning speed adjustment range is 100-420mm/s
±10%.
Buttons for coarse feed backward/forward fast.
•
Fast advance/retract movement of the coarse feed.
•
Speed is 900µm/s.
Buttons for coarse feed backward/forward slow
•
Slow advance/retract movement of coarse feed.
•
Speed is 300µm/s.