5
INSTRUCTIONS
1.
Remove the equipment from the shipping carton carefully. Save the shipping
carton and all packing material until proper operation of the equipment is
confirmed. Immediately notify the carrier if there is any visible damage to the
carton or its contents. Failure to do so may limit your ability to obtain
compensation for damage.
NOTE:
The inner chamber of your instrument has been treated with an oil-based
protectant. Please clean the chamber thoroughly with alcohol before filling to avoid
the contamination of paraffin.
2.
Connect the power cable to the unit and into a properly grounded 120V outlet.
NOTE:
If your unit was shipped outside of the U.S., check the label on the rear of
the instrument for the electrical requirements.
3.
Switch the
ON/OFF
switch at the rear of the unit to the
ON
position.
4.
Immediately upon turning the unit on, the
SET TEMP
and
TIME/TEMP
windows
will display
8888
for approximately 4 seconds. The unit will then display
PASS
or
FAIL
in the
TEMP/TIME
window for 4 seconds. If
PASS
appears,
06:00
will
appear in the red LED window and flash until set, while the
SET TEMP
window will
display the factory preset temperature of
55°C
.
NOTE:
If
FAIL
appears in the
TEMP/TIME
window, the
SET TEMP
LED will be
blank and the beeper will continue to sound. Press any switch to discontinue the
beeping. Although the water bath may appear to function normally, the
FAIL
message indicates that the software has found a problem with the unit, and the
unit should be returned for service. Phone our customer service department
immediately for further instructions.
5.
Press the
ON
key to initiate the paraffin reservoir heaters. The small red LED
indicator will light up.
6.
You may now set the clock by pressing the
SET TIME FAST
or
SET TIME SLOW
keys. Press the
TEMP
key to the left of the
TEMP/TIME
display if you prefer
continuous monitoring of the bath temperature.
7.
To increase or decrease the preset reservoir temperature, press and hold the
increment (up arrow) key or the decrement (down arrow) key until the desired