INSTRUCTIONS 

 
 
 
 

1. 

Remove the equipment from the shipping carton carefully. Save the shipping 
carton and all packing material until proper operation of the equipment is 
confirmed. Immediately notify the carrier if there is any visible damage to the 
carton or its contents. Failure to do so may limit your ability to obtain 
compensation for damage. 

 

NOTE:

 The inner chamber of your instrument has been treated with an oil-based 

protectant. Please clean the chamber thoroughly with alcohol before filling to avoid 
the contamination of paraffin. 

 
2. 

Connect the power cable to the unit and into a properly grounded 120V outlet. 

 

NOTE:

 If your unit was shipped outside of the U.S., check the label on the rear of 

the instrument for the electrical requirements. 

 
3. 

Switch the 

ON/OFF

 switch at the rear of the unit to the 

ON

 position. 

 
4. 

Immediately upon turning the unit on, the 

SET TEMP

 and 

TIME/TEMP

 windows 

will display

 

8888

 for approximately 4 seconds. The unit will then display 

PASS

 or 

FAIL

 in the 

TEMP/TIME

 window for 4 seconds. If 

PASS

 appears, 

06:00

 will 

appear in the red LED window and flash until set, while the 

SET TEMP

 window will 

display the factory preset temperature of 

55°C

 

NOTE:

 If 

FAIL

 appears in the 

TEMP/TIME

 window, the 

SET TEMP

 LED will be 

blank and the beeper will continue to sound. Press any switch to discontinue the 
beeping. Although the water bath may appear to function normally, the 

FAIL

 

message indicates that the software has found a problem with the unit, and the 
unit should be returned for service. Phone our customer service department 
immediately for further instructions. 

 
5. 

Press the 

ON

 key to initiate the paraffin reservoir  heaters. The small red LED 

indicator will light up. 

 
6. 

You may now set the clock by pressing the 

SET TIME FAST

 or 

SET TIME SLOW

 

keys. Press the 

TEMP

 key to the left of the 

TEMP/TIME

 display if you prefer 

continuous monitoring of the bath temperature. 

 
7. 

To increase or decrease the preset reservoir temperature, press and hold the 
increment (up arrow) key or the decrement (down arrow) key until the desired 

Summary of Contents for PD-120

Page 1: ...hly before proceeding with unpacking and installation TRIANGLE BIOMEDICAL SCIENCES INC 3014 CROASDAILE DRIVE DURHAM NORTH CAROLINA 27705 2507 USA 919 384 9393 FAX 919 384 9595 E MAIL tbs trianglebiome...

Page 2: ...E TABLE OF CONTENTS Topic Page Warranty card Insert A Introduction 1 Symbols and conventions 2 Overview and specifications 3 Instructions 5 Precautions and maintenance 8 Regulatory Compliance 8 Error...

Page 3: ...oprocessor controlled to maintain 1 C temperature accuracy within the paraffin reservoir A sensor positioned on the outside cylinder wall detects temperature changes precipitated by the introduction o...

Page 4: ...2 SYMBOLS AND CONVENTIONS Important Notice Refer to documentation before operation Hot Surface...

Page 5: ...3 OVERVIEW AND SPECIFICATIONS TBS Paraffin Dispenser PD...

Page 6: ...46 1mm Width 14 355 6mm Depth 16 406 4mm Voltage 120V AC 10 50 60Hz single phase Current 4 5 amps Power 540 watts Default temperature setting 55 C Maximum temperature 60 C Heat output 1885 BTU hr typi...

Page 7: ...mately 4 seconds The unit will then display PASS or FAIL in the TEMP TIME window for 4 seconds If PASS appears 06 00 will appear in the red LED window and flash until set while the SET TEMP window wil...

Page 8: ...display in the TIME mode for continuous monitoring as a clock or press the TEMP key to view the actual paraffin temperature It is not necessary to unplug the paraffin dispenser while not in use for s...

Page 9: ...ng the unit off TBS recommends that the unit be connected to a surge protector or some form of uninterrupted power source to ensure against such disruptions Occasionally check the filter located in th...

Page 10: ...gorous international quality and safety standards and has been found to be in compliance with the following regulations and specifications UL 61010A CSA 1010 1 CE EN IEC EN61010 1 Safety EN61000 3 2 H...

Page 11: ...r sounds an alarm and the SET TIME LED display flashes an error code The error code indicates one of the specific problems as described below Error code Problem E0 General system failure E1 Sensor 1 T...

Page 12: ...pense wax Defective spigot heater Defective solid state relay K4 Check resistance value for approximately 1200 at P3 pin 2 and P4 pin 6 Replace controller board relay call TBS Paraffin completely melt...

Page 13: ...r calibration procedure E2 error code display reads HI but there is no indication of overheating Defective band temperature sensor P2 disconnected Controller board failure Call TBS for replacement Rec...

Page 14: ...angle Biomedical Sciences Inc 3014 Croasdaile Drive Durham NC 27705 U S A Sales and product information sales trianglebiomedical com Customer service questions service trianglebiomedical com All other...

Reviews: