TAU STERIL Tau Clave 3000 Instructions For Use Manual Download Page 9

 

TAU STERIL s.n.c 

Via Gorizia, 6/a 

Fino Mornasco ( CO ) 

 

 

Instructions for Use 

TAU CLAVE 3000 Fractionated Vacuum 

Data: 30.03.15 

Pag. 9 di 14 

Rev. 3 

N° 1 

 

 

This document may not be reproduced in partial form without the authorization of TAU STERIL. 

 

 

 

 
 
There are also other alarms that are related to the components of the electronic board; they are: 
 
 

CODICE  ALLARME 

INDICAZIONI 

A 20 

Data loss * 

A 21 

Watchdog malfunction * 

A 22 

Defective sensors * 

A 23 

Batteries Clock * 

A 24 

Software* 

 

* = Repeat cycle, the material to be sterilized is not sterile. If the problem persists, contact the manufacturer. 

 

5.4  Instruction Manual 

 

1

 

Clean the material to be sterilized. 

2

 

Turn on the main switch. 

3

 

 Load the cargo tank (capacity 2.5 machine completely empty), until the message "STOP" on the display time.                   
It occurs whenever the alarm Low H20 LEVEL add 1.5 liters of demineralized water. If you go over, it arises "STOP" that 
is  tank full load, then pull the tube out of the oil cargo. 

4

 

Whenever the alarm A02 occurs, download the recovery tank (better not wait for the alarm because nothing prevents you 
download it at any time) with the appropriate tube. 

5

 

Load the media on the trays provided (without overlapping the pieces) and pull it into the sterilization chamber. 

6

 

Select the program: to do this press the select button  

A single press of the button does not change the program, but it serves the display shows the temperature of the temperature 
program:  After five seconds the display temperature resumes its normal content. Within five seconds if the select button is 
pressed again, the program is changed (indicated by LED).  If the selected program is special, the contents of the display 
temperature will flash to indicate that the sterilization temperature can be chosen between 105 ° -  134 ° with the + - 
temperature. The same applies to the time with the + - time. 
These variations can be carried out simultaneously in both the time for which the temperature 
7

 

Close the door after you have done so. 

8

 

Press Start: starts the selected cycle. 

9      Keep area within the opening of the door, to avoid (the opening of the door in hot machine) burns or other hazards:   
 

WARNING: DURING USE USE GLOVES. 

 

NOTE: At any time (ex: During sterilization or during DRY = dry) you can break the cycle by pressing the "STOP", a buzzer 
will sound and the beeper in the presence of pressure will be downloaded automatically.

 

 

5.5

 

 

Description cycle TAU CLAVE 3000 Fractionated Vacuum 

 

The phases of the cycle take place as follows: 
At the start of the cycle we have the first vacuum phase (Cycle Unwrapped - Wrapped - Porous). 
Follow the steps of fractionated vacuum with placing water cycle Hollow - Special). 
This system is suitable for the sterilization of hollow bodies such as turbines, etc. autoclavable cannulas. etc. 
The sterilization step takes place while maintaining the machine at a constant pressure. Then follows the drying phase is also 
composed of different phases of vacuum and ventilation. A sterilization cycle completed (END appears on the display 
"Temperature"). 
At this point part of the step of which consists in ventilation-condensation ventilate at intervals of 2 minutes (through  the 
pump-vacuum) the interior of the sterilization chamber of the autoclave until the temperature read on the display "temperature" 
does not drop a degree below 70 degrees C °. This allows to obtain materials perfectly dry. The machine can still be open at 
any time by opening the door (see movement opening - closing the door 2.1 p. 3).

 

 

Summary of Contents for Tau Clave 3000

Page 1: ... TAU CLAVE 3000 Fractionated Vacuum Data 30 03 15 Pag 1 di 14 Rev 3 N 1 1 This document may not be reproduced in partial form without the authorization of TAU STERIL Instructions for Use Tau Clave 3000 Fractionated Vacuum 0426 DIRECTIVE 93 42 EEC and S M I MEDICAL DEVICES ...

Page 2: ...ning and closure 4 3 Installation 5 4 Characteristics 6 4 1 Technical data 6 4 2 Environmental working conditions 6 4 3 Recommended times and characteristics of programmes 6 5 Operating modes 7 5 1 General information 7 5 2 Visualizations 7 5 3 Alarms visualization modes 8 5 3 1 Alarms and corresponding summary codes 8 5 4 Operating instructions 9 5 5 Description of Tau Clave 3000 Fr Vacuum cycle ...

Page 3: ... CEI EN ISO 14971 The autoclave Tau 3000 Fractionated Vacuum complies with RoHS 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment 2 2 Expected use of materials to sterilize The equipment is used to sterilize surgical instruments in the medical paramedical sector The materials ...

Page 4: ...O Instructions for Use TAU CLAVE 3000 Fractionated Vacuum Data 30 03 15 Pag 4 di 14 Rev 3 N 1 4 This document may not be reproduced in partial form without the authorization of TAU STERIL 2 4 Door opening and closure OPENING MOVEMENT CLOSING MOVEMENT ...

Page 5: ... under the support legs front it is important not to touch the shims On the rear side of the machine is located in a pipe for the steam to exit IMPORTANT The tube must be positioned as shown in Figure 2 Fig 3 Fig 4 Fill the bottle with 2 5 liters of demineralized water then close the bottle with the cap see Fig 4 Introduce the cannula orifice load autoclave Unscrew the tap until the water begins t...

Page 6: ... with phases of ventilation To select the cycle BOWIE DICK TEST as follows Place the program cycle HOLLOW before viewing 5 sec Of the cycle has finished press the button decreases the time Automatically displays the cycle 134 3 5 of BOWIE DICK TEST PROGRAM FRACTIONATED VACUUM STERILIZATION DRYING AND VENTILATION VACUUM FINAL UNWRAPPED Instruments unwrapped YES 134 4 Minutes 2 11 Bar 15 Minutes YES...

Page 7: ...se the time in the program Special 6 INCREASE TIME to increase the time in the program Special 5 2 Views 1 PRESSURE It indicates the pressure in bar 2 TEMPERATURE Shows the temperature of steam in the chamber or by pressing the right button TEMP shows the temperature of the heating element Also pressing the left button TEMP you can see the selected temperature cycle 3 TIME Indicating the steriliza...

Page 8: ...s alarm occurs during the cycle the latter is interrupted and the pressure discharged 5 3 1 Alarms and codes Summary CODE ALARM INDICATIONS A00 Door open occurs if the start of the cycle the door is open Disappears by pressing any key or by closing the door A01 Low water occurs if at the start of the cycle the water level in the cargo tank is insufficient Disappears by pressing any key or by reset...

Page 9: ...pecial the contents of the display temperature will flash to indicate that the sterilization temperature can be chosen between 105 134 with the temperature The same applies to the time with the time These variations can be carried out simultaneously in both the time for which the temperature 7 Close the door after you have done so 8 Press Start starts the selected cycle 9 Keep area within the open...

Page 10: ...sufficient ventilation Avoid placing the autoclave near sources of steam or splashing with water that may damage the internal electronics Do not install in places with poor ventilation Do not place it near heat sources 7 Maintenance and Cleaning Each type of maintenance must be carried out with power wire off and the machine completely cold WARNING Do not modify the equipment without the authoriza...

Page 11: ...e at the side of the probe inside the sterilization chamber see fig 5 Fig 5 Dismantling With the cold machine turn from left to right the steel tube with the help of pliers if necessary paying attention to the probe closer Remove the piece from the chamber unscrew the filter clean it with vinegar or if too dirty replace it with a new one If you want you can avoid disassembling the tube with filter...

Page 12: ...maximum load 14 Floating common load connection 15 Micro switch door 16 Electric valve N A Normally open 2 Way 24V DC 17 Output optional 24V cc 18 Electric valve N C 2 Way 24V DC 19 Electric valve N C 2 Way 24V DC 20 230 V power supply feeding the vacuum pump and the electric valve nc 2 Vie 21 Bimetallic security 22 Power supply 230 V 50 Hz 23 Resistance 230V 1250 Watts 24 switch 25 Fuse 6 3x32 qu...

Page 13: ...and electrical equipment or put them out of order It s forbidden the operator to carry out repairs of any kind During cleaning and or maintenance it is absolutely essential to ensure that the equipment is disconnected from the power supply and it is completely cooled In the case of sterilization of surgical instruments eg scalpel or pointed instruments we recommend the use of safety shoes Periodic...

Page 14: ...Rev 3 N 1 14 This document may not be reproduced in partial form without the authorization of TAU STERIL 10 2 Explanation of Symbols WARNING READ THE INSTRUCTIONS WARNING DANGER BURNS CAUTION RISK OF ELECTRIC SHOCK WARNING USE GLOVES FOR HAND PROTECTION PRODUCT NOT COMPARABLE TO MUNICIPAL WASTE FOLLOW THE RULES FOR MUNICIPAL DISPOSAL ...