Instructions for Use
TAU CLAVE 3000 Fractionated Vacuum
Data: 30.03.15
Pag. 9 di 14
Rev. 3
N° 1
9
This document may not be reproduced in partial form without the authorization of TAU STERIL.
There are also other alarms that are related to the components of the electronic board; they are:
CODICE ALLARME
INDICAZIONI
A 20
Data loss *
A 21
Watchdog malfunction *
A 22
Defective sensors *
A 23
Batteries Clock *
A 24
Software*
* = Repeat cycle, the material to be sterilized is not sterile. If the problem persists, contact the manufacturer.
5.4 Instruction Manual
1
Clean the material to be sterilized.
2
Turn on the main switch.
3
Load the cargo tank (capacity 2.5 machine completely empty), until the message "STOP" on the display time.
It occurs whenever the alarm Low H20 LEVEL add 1.5 liters of demineralized water. If you go over, it arises "STOP" that
is tank full load, then pull the tube out of the oil cargo.
4
Whenever the alarm A02 occurs, download the recovery tank (better not wait for the alarm because nothing prevents you
download it at any time) with the appropriate tube.
5
Load the media on the trays provided (without overlapping the pieces) and pull it into the sterilization chamber.
6
Select the program: to do this press the select button
A single press of the button does not change the program, but it serves the display shows the temperature of the temperature
program: After five seconds the display temperature resumes its normal content. Within five seconds if the select button is
pressed again, the program is changed (indicated by LED). If the selected program is special, the contents of the display
temperature will flash to indicate that the sterilization temperature can be chosen between 105 ° - 134 ° with the + -
temperature. The same applies to the time with the + - time.
These variations can be carried out simultaneously in both the time for which the temperature
7
Close the door after you have done so.
8
Press Start: starts the selected cycle.
9 Keep area within the opening of the door, to avoid (the opening of the door in hot machine) burns or other hazards:
WARNING: DURING USE USE GLOVES.
NOTE: At any time (ex: During sterilization or during DRY = dry) you can break the cycle by pressing the "STOP", a buzzer
will sound and the beeper in the presence of pressure will be downloaded automatically.
5.5
Description cycle TAU CLAVE 3000 Fractionated Vacuum
The phases of the cycle take place as follows:
At the start of the cycle we have the first vacuum phase (Cycle Unwrapped - Wrapped - Porous).
Follow the steps of fractionated vacuum with placing water cycle Hollow - Special).
This system is suitable for the sterilization of hollow bodies such as turbines, etc. autoclavable cannulas. etc.
The sterilization step takes place while maintaining the machine at a constant pressure. Then follows the drying phase is also
composed of different phases of vacuum and ventilation. A sterilization cycle completed (END appears on the display
"Temperature").
At this point part of the step of which consists in ventilation-condensation ventilate at intervals of 2 minutes (through the
pump-vacuum) the interior of the sterilization chamber of the autoclave until the temperature read on the display "temperature"
does not drop a degree below 70 degrees C °. This allows to obtain materials perfectly dry. The machine can still be open at
any time by opening the door (see movement opening - closing the door 2.1 p. 3).