Instructions for Use
TAU CLAVE 3000 Fractionated Vacuum
Data: 30.03.15
Pag. 10 di 14
Rev. 3
N° 1
10
This document may not be reproduced in partial form without the authorization of TAU STERIL.
6 Conditions for the Transport / Handling and Storage
The autoclave is supplied with packaging that requires the usual precautions for transport fragile materials, as indicated on the
packaging boxes himself. When returning for repair, use the original packaging. When storing, place the product in areas
without humidity to protect the electrical components.
The device is packed in the following manner:
Two caps of polyethylene (above and below) protected by a box around the volcano in triple wave.
Do not lift the machine with violent jerks and not flip.
The device and the packaging are delicate, handle with care. Carry no shock, impact. Store in a dry and protected. The
packaging should be stored.
Using a different packaging, it could cause damage to the product during shipment.
The appliance must be extracted with the help of two people
- Remove the top cap.
- Extract the machine , taking care to keep it always in a horizontal position.
- Place the unit on the worktop.
- The device must be installed inside a laboratory accessible only to authorized personnel.
- Place the unit on a flat, horizontal surface.
- Leave a space of at least 20 cm around the apparatus for sufficient ventilation.
- Avoid placing the autoclave near sources of steam or splashing with water that may damage the internal electronics.
- Do not install in places with poor ventilation.
- Do not place it near heat sources.
7 Maintenance and Cleaning
Each type of maintenance must be carried out with power wire off, and the machine completely cold.
WARNING: Do not modify the equipment without the authorization of the manufacturer.
In case of malfunction, contact the manufacturer.
For repairs to make the sterilizer manufacturer in its original packaging.
Cleaning the sterilization chamber (to be repeated on a daily basis):
•
A cold machine using a sponge with vinegar to remove limestone from the chamber
•
Wash the sponge with demineralized water and clean the chamber itself (several times)
•
Pass a clean cloth with some isopropyl alcohol (90%) within the sterilization chamber.
Maintenance of the bacterial filter:
•
Replace the filter every 6 months
•
If dirt (sucks outside air during the sterilization cycle) to replace the first.
A cold machine monthly check that the door seal is accurate in the hard (steel) that contains: if withdrawn replace with a new
one. We recommend replacing the gasket still every six months or after 200 sterilization cycles. Clean the seal on a daily basis
(the machine is cold) with alcohol repositioning it in its proper housing.
Check monthly (or even if you move the device) the electric power cable is intact and in good condition.
In case of malfunction contact the manufacturer.
For repairs, it is recommended to make the autoclave to the manufacturer, using the original packaging