Control Board 
 

1= Thermostat 
2= Thermostat bulb 
3= Functioning bulb 
4= Timer 

 

SAFETY NORMS AND PRECAUTIONS 

 

During  usage  and  while  extracting  your  instruments  from  the  oven,  wear 
protective  gloves  to  avoid  burnings  (outside  the  plate  warns: 

BEWARE: 

BURNING DANGER!) 

 

TAKE CARE WHEN EXTRACTING THE INSTRUMENTS FROM THE OVEN 
DANGER OF BURNING, WEAR PROTECTIVE GLOVES. 

 

This hot-air oven must be used only by authorised and specialised 

personnel who are familiar with sterilisation techniques, times and 

temperatures. 

 
This  device  must  be  used  only  for  the  purpose  it  has  been  meant  for: 
sterilisation of surgical instruments by surgeons or paramedical staff.  

It is necessary

 to take any possible precautions to avoid 

improper

 use by 

unauthorised personnel

. It is forbidden to modify or alter in any way the 

thermic and electric safety protections of the device or put them out of use. 
It is also forbidden to try to repair the machine on your own. 
When  servicing  or  cleaning  the  machine, 

make  sure

  it  is  unplugged  and 

completely cooled down.  
In  case  of  sterilisation  of  surgical  instruments  (scalpels  or  pointed 
instruments)  wear protection shoes.  
Periodically check that t the supply cable is not damaged. 
 

SYMBOLS 

 

IMPORTANT: READ INSTRUCTIONS FOR USE   

 

 

BEWARE: BURNING DANGER  

 

 
 

Summary of Contents for MINI AUTOMATIC

Page 1: ...TAU STERIL s n c Via Gorizia 6 a Fino Mornasco CO TAU STERIL MINI AUTOMATIC INSTRUCTIONS FOR USE CE 0426 ...

Page 2: ...een manufactured according to the EC norms The usage of this hot air oven is limited to those functions for which it has been projected and realised The Manufacturer declines any responsibility on risks or dangers due to improper usage by operators Guarantee lasts 12 months starting form the invoice date INSTALLATION Verify that transport and or stocking have respected acceptable humidity conditio...

Page 3: ... values documented and declared by the Manufacturer of the material and or instrument to be sterilised Suggested temperatures by hot air sterilizers range from 150 C to 180 C Important the operator must respect the suggested sterilisation temperatures The maximum adjustable temperature is 200 C The suggested sterilisation times range from a minimum of 30 minutes to a maximum of 45 60 minutes The m...

Page 4: ...tion chamber DAILY with a cloth moistened with denaturated alcohol 90 Not do spray water do not let water penetrate the chamber to avoid electric shocks Check once a month that the supply cable is in good conditions In case of bad functioning contact the Manufacturer In case of repair we recommend that you send back the device to the Manufacturer with its original package Check periodically yearly...

Page 5: ...ed only for the purpose it has been meant for sterilisation of surgical instruments by surgeons or paramedical staff It is necessary to take any possible precautions to avoid improper use by unauthorised personnel It is forbidden to modify or alter in any way the thermic and electric safety protections of the device or put them out of use It is also forbidden to try to repair the machine on your o...

Page 6: ...6 BEWARE DANGER OF ELECTRICAL SHOCK BEWARE WEAR PROTECTIVE GLOVES AEE WASTE ACCORDING TO LOCAL RULES ...

Reviews: