WARNING 

 
Tau Steril Mini Automatic is a hot-air steriliser: do not put any humid objects 
inside the sterilisation chamber to avoid electric shocks. 
 

INTRODUCTION 

 

Norms 

Tau Steril Mini Automatic complies with the essential features indicated in 
the Directive 93/42/EC dated June 14

th

 1993, class II A and with the 

following technical norms: 
 
EN 61010-1 
EN 61010-2-040 
EN61326-1 
UNI CEI EN ISO 14971

 

 

Materials which can be sterilised: 

The oven is meant for the sterilisation of surgical instruments by surgeons or 
paramedical  staff.  The  materials  to  be  sterilised  must  stand  up  to  the  high 
working  temperatures,  i.e.  from  0  °C  to  200  °C. 

Do  not  insert  any  toxic 

material 

inside, 

the 

Manufacturer 

Tau 

Steril 

declines 

any 

responsibility to this respect. 

 

MANUFACTURER’S RESPONSIBILITIES 

 
This device has been manufactured according to the EC norms. The usage of 
this hot air oven is limited to those functions for which it has been projected 
and  realised.  The  Manufacturer  declines  any  responsibility  on  risks  or 
dangers due to improper usage by operators. 
Guarantee lasts 12 months starting form the invoice date. 
 

INSTALLATION 

 
Verify  that  transport  and/or  stocking  have  respected  acceptable  humidity 
conditions in order not to affect the safety of the sterilizer. Stock and install 
the  device  in  a  humidity  safe  place  (Tau  Steril  declines  any  responsibilities 
concerning  the  safety  of  the  device  if  this  should  be  kept  in  an  unsuitable 
way). 
Remove  the  device  from  its  packaging  with  extreme  care  and  check  its 
integrity.  
Verify that electric supply indicated on the label on the back corresponds to 
to  the  one  available  in  the  installation  area  (max  variation  of  electrical 
supply 

±

 10%). 

Verify that the main electric plant has an earth, set the oven on a horizontal 
surface to grant plenty of ventilation.  

Summary of Contents for MINI AUTOMATIC

Page 1: ...TAU STERIL s n c Via Gorizia 6 a Fino Mornasco CO TAU STERIL MINI AUTOMATIC INSTRUCTIONS FOR USE CE 0426 ...

Page 2: ...een manufactured according to the EC norms The usage of this hot air oven is limited to those functions for which it has been projected and realised The Manufacturer declines any responsibility on risks or dangers due to improper usage by operators Guarantee lasts 12 months starting form the invoice date INSTALLATION Verify that transport and or stocking have respected acceptable humidity conditio...

Page 3: ... values documented and declared by the Manufacturer of the material and or instrument to be sterilised Suggested temperatures by hot air sterilizers range from 150 C to 180 C Important the operator must respect the suggested sterilisation temperatures The maximum adjustable temperature is 200 C The suggested sterilisation times range from a minimum of 30 minutes to a maximum of 45 60 minutes The m...

Page 4: ...tion chamber DAILY with a cloth moistened with denaturated alcohol 90 Not do spray water do not let water penetrate the chamber to avoid electric shocks Check once a month that the supply cable is in good conditions In case of bad functioning contact the Manufacturer In case of repair we recommend that you send back the device to the Manufacturer with its original package Check periodically yearly...

Page 5: ...ed only for the purpose it has been meant for sterilisation of surgical instruments by surgeons or paramedical staff It is necessary to take any possible precautions to avoid improper use by unauthorised personnel It is forbidden to modify or alter in any way the thermic and electric safety protections of the device or put them out of use It is also forbidden to try to repair the machine on your o...

Page 6: ...6 BEWARE DANGER OF ELECTRICAL SHOCK BEWARE WEAR PROTECTIVE GLOVES AEE WASTE ACCORDING TO LOCAL RULES ...

Reviews: