3
Do not position it less than 10 cm from walls or other devices to allow proper
ventilation around the machine. Do not obstruct the correct opening of the
door with any object.
Plug in to the wall socket and position the main switch to ON; make sure
that the device is on.
Tau Steril declines any responsibilities if these instructions are not observed.
CHARACTERISTICS
TECHNICAL FEATURES
Weight 3,5 kg
External dimensions: cm 31 x h cm 18 x cm 18
Inner dimensions: cm 24.5 x h cm 5,5 x cm 11,5
Absorption: 270 W
Supply: 230 V 50 Hz.
Operating conditions
Room Temperature: 5° to 40 ° C
Humidity: (at maximum temperature of 31°C) 80%
Humidity (at maximum temperature of 40°C) 50%
Suggested times and temperatures
The sterilisation temperatures must correspond to the specific values
documented and declared by the Manufacturer of the material and /or
instrument to be sterilised. Suggested temperatures by hot air sterilizers
range from 150°C to 180°C.
Important: the operator must respect the
suggested sterilisation temperatures.
The maximum adjustable temperature is 200°C
.
The suggested sterilisation times range from a minimum of 30
minutes to a maximum of 45-60 minutes.
The maximum adjustable time is 120 minutes.
HOW TO USE
1.
Carefully clean and dry the instruments to be sterilised.
2.
Plug the device in to the main
3.
To set the right temperature
: rotate the thermostat knob to reach the
desired temperature
4.
Rotate the timer know clockwise to set the desired sterilisation time
(max. 120 min)
5.
At the end, the two lights will confirm that the machine is running (green
light = functioning, yellow light = thermostat). Before stabilising on the
set temperature the yellow light will flash to indicate the rising
temperature inside the chamber.