![TASKING iSYSTEM RH850 User Manual Download Page 12](http://html1.mh-extra.com/html/tasking/isystem-rh850/isystem-rh850_user-manual_3654466012.webp)
Hardware Setup and Configuration
For detailed visual presentation of the hardware setup and configuration, refer to
. Use short link isystem.com/start.
1. Connect the power supply cable. BlueBox should be switched off.
2. First connect via USB. Later you can configure TCP/IP connection to work remotely.
3. Connect the Grounding wire to the BlueBox and the Target.
If the Grounding wire is not connected, the ground potential difference between the
BlueBox and the Target can exceed well over 1000V even before any of the devices
are powered up. This voltage difference is discharged over the BlueBox and the
Target, leading to the possible destruction of electronic components.
4. Connect FNet cable of Active Probe to the BlueBlux FNet port.
Although it looks similar to the HDMI interface, the FNet Port is
not compatible with
HDMI or any HDMI accessories
.
Connecting iSYSTEM hardware to HDMI accessories will damage the hardware and
will render the iSYSTEM hardware warranty void.
5. Power ON the hardware in the following order:
a. BlueBox
b. Target
6. Install winIDEA and create a new workspace.
7. Configure Debug channel modes via
Hardware / CPU Options / SoC
.
For troubleshooting refer to
- kb.isystem.com.