Specifications

Names and connection methods for signal lines

Weighing Mode

BMI Mode

6. RS-232C Communication Specifications

Communication standard

EIA RS-232C

Communication method

Asynchronous method

Signal speed

2,400 baud

Data bit length

7 bit

Parity

EVEN

Stop bit

1 bit

Terminal no.

Signal name

1
2

RXD

3

TXD

4
5

GND

6
7
8
9

An RS-232C connector (D sub 9 pin male) is attached to the side of the display box.

When connecting the display box with a personal computer, etc., please use an RS-232C cross cable.

When receiving 4 bytes of data as indicated in 1-1 while using the scale in weighing mode, weight data will be output in 14 bytes, as indicated in 1-2.
* While the scale is used with lb (pound) measurement units, the weight data will be output in lbs (pounds).

While the scale is used in BMI mode, the output data will vary depending on the form of the received data.

When receiving 4 bytes of data, as indicated in 1-1, the scale will calculate your BMI using the currently input height figure, and 26 bytes
of data will be output, as indicated in 2-1 or 2-2, depending on the measurement unit.

When receiving 11 bytes of data, including measurement units and height figures, in the form indicated in 1-2 or 1-3, the scale will calcu-
late your BMI using these figures, and 26 bytes of data will be output, as indicated in 2-1 or 2-2, depending on the measurement unit.

Details of Weighing Mode Transmission Data

Byte nos. 1-2

Byte nos. 5-10

Byte nos. 3-4

Byte nos. 11-12
Byte nos. 13-14

Header

Separator

Weight data

Measurement units

Terminator

2 bytes fixed

2 bytes fixed
6 bytes fixed
2 bytes fixed
2 bytes fixed

SD: Unstable  S_: Stable

SI: Overload, etc.

2 spaces

Right-aligned

Kg

:

kg

Lb:lb

CR+LF

1-1 Reception data

D R

CR LF

1

2

3

4

1-2 Transmission data

S D

7 5

.

0 k g

CR LF

1

2

3

4

5

6

7

8

9

10 11 12 13 14

1-1 Reception data

D R

CR LF

1

2

3

4

1-2 Reception data (height data when using 

kg

)

*Please input 90-99cm height data as 090-099.

D H k g

1 8 0

CR LF

1

2

3

4

5

6

7

8

9

10 11

1-3 Reception data (height data when using lb)

*  Height data in ft. (feet) and in. (inches) will appear as follows.

5 feet, 11.0 inches will appear as 5.110.

* When INCH is 1 figure less than 10 inch, please put in 0 and input by 2 figures.

e.g. 2inch 

02 inch

D H L b 5

.

1 1 0

CR LF

1

2

3

4

5

6

7

8

9

10 11

2-1 Transmission data (

kg

/cm)

S D

7 5

.

0 k g

,

1 8 0

,

2 3

.

1

CR LF

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26

2-1 Transmission data (lb/ft. in)

S D

1 6 5

.

4 L b

,

5

.

1 1 0

,

2 3

.

1

CR LF

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26

Details of BMI mode transmission data

Byte nos. 1-2

Byte nos. 5-10

Byte nos. 3-4

Byte nos. 11-12

Byte nos. 13

Byte nos. 14-18

Byte nos. 19

Byte nos. 20-24
Byte nos. 25-26

Header

Separator

Weight data

Measurement units

Separator

Height data

Separator

BMI data

Terminator

2 bytes fixed

2 bytes fixed
6 bytes fixed
2 bytes fixed

1 byte fixed

5 bytes fixed

1 byte fixed

5 bytes fixed
2 bytes fixed

SD: Unstable  S_: Stable

SI: Overload, etc.

2 spaces

Right-aligned

Kg

:

kg

Lb:lb

1 comma

Right-aligned

1 comma

Right-aligned

CR+LF

1

6

9

5

GB

9

8

Summary of Contents for WB-100

Page 1: ...pper 5Digits and Lower 4Digits LCD Height of numerals 25mm 1in RS 232C D sub 9pins Male Connector AC Adapter included Center Minus DC9V 300mA LR6 AA Alkaline Battery 6 not included 0 3W max 0ºC 35ºC 32ºF 95ºF 5 1kg 11 3lb approximately 100 hours of continuous use when using LR6 AA Alkaline battery 270kg 0 1kg increments 600lb 0 2lb increments 200kg 0 1kg increments 440lb 0 2lb increments Weight Me...

Page 2: ...duct pack it in its original carton Warnings Various symbols are used in this instruction manual as well as on the product in order to prevent injury to personnel and damage to property It is advisable to familiarize yourself with these terms prior to reading the rest of this booklet WARNING Death or serious injury could occur if these precautions are not followed or the product is misused WARNING...

Page 3: ...splay box Please screw the wall hanging screws into the wall leave the screws protruding from the wall by about 5 mm 0 2in Please insert the screw heads in the holes on the back of the display box Please pull down the display box lightly with your hands If the unit is not installed properly there is a risk that the display box will fall down Please avoid installing the display box in a place where...

Page 4: ...BATTERY kg lb cm ft in NET HEIGHT BMI BMI MODE STABILIZED STEP ON kg NET STABILIZED STEP ON kg NET STABILIZED STEP ON kg NET STABILIZED STEP ON ON ZERO OFF d d d d d d d kg NET STABILIZED STEP ON kg NET STABILIZED STEP ON kg NET STABILIZED STEP ON kg NET STABILIZED STEP ON LO BATTERY kg lb cm ft in NET HEIGHT BMI BMI MODE STABILIZED STEP ON ON ZERO d d 5 Operation 1 Standard Weighing Procedures 2 ...

Page 5: ...s 5 10 Byte nos 3 4 Byte nos 11 12 Byte nos 13 14 Header Separator Weight data Measurement units Terminator 2 bytes fixed 2 bytes fixed 6 bytes fixed 2 bytes fixed 2 bytes fixed SD Unstable S_ Stable SI Overload etc 2 spaces Right aligned Kg kg Lb lb CR LF 1 1 Reception data D R CR LF 1 2 3 4 1 2 Transmission data S D 7 5 0 k g CR LF 1 2 3 4 5 6 7 8 9 10 11 12 13 14 1 1 Reception data D R CR LF 1 ...

Page 6: ...GB GB 11 10 ...

Reviews: