A RS-232C connector (D-sub 9-pin female) is attached to the side of the control unit.
The RS-232C extension cable (straight cable) is required for connection to a computer, etc. Some commercially
available cables cannot be connected to this product. The connection cable to the PC must be used less than 3m.
When connecting a computer or peripherals, be sure that compliance with IEC60601-1 (EN60601-1) is
maintained.
If the power is supplied in compliance with IEC60950 (EN60950), power must be supplied from a medical
isolating transformer. Keep at a minimum distance of 1.5m between each item of equipment during operation.
Note
Output Data Format
*The RS232C port is attached to left of the control unit.
24
en
If necessa
ry
Specifications
Signal Line Name and Connection Method
Connection Example
This explanation is for the interface to output the readings from PW-650MA to external equipment (e.g. PC) as a
RS-232C-compliant signal.
Communication standard
EIA RS-232C
Communication method
Asynchronous
Signal speed
9,600bps
Data bit length
8bit
Parity
NONE
Stop bit
1bit
Terminator
CR+LF
Flow control
None
Terminal no.
Signal name
1
2
TXD
3
PW-650MA
PC, etc.
4
5
GND
6
7
8
9