Tando TR-71S User Manual Download Page 12

22

Collecting Recorded Data

With our exclusive software "Thermo Recorder for Windows" it is possible to 
download all of the recorded data from TR-71S and TR-72S as well as carry 
out the following functions in managing the data.

-  Display 8 channels of data in colorful graphs
-  Control Display ON / OFF for each channel
-  Zoom in and out of graph data
-  Display data for two points on graph and differences in 

measurements for those points

-  Select calculation range and calculate highest, lowest and average 

values for that range

-  Select and set precise recording start date and time
-  Change channel names
-  Change channel order
-  Merge channels
-  Delete channels
-  Change graph background color, line color and line width.
-  Create data tables
-  Print out graphs and tables
-  Create and export data in Text file format
-  Save data files

 For more details about these functions and operation of the software , see 

Thermo Recorder for Windows User’s Manual or the "HELP" menu of the 
software.

Temperature Sensors (for TR-7 Series)

 

Possible Measurement Range:  -40 to 110

 

Sensor Temperature Durability:  -50 to 115

 

Measurement Accuracy:  Average

±

0.3

℃(

-20 to 80

℃)

 

 

Average

±

0.5

℃(

-40 to -20

80 to 110

℃)

Materials 

Thermistor 

TPE resin-shielded sensor 

TPE resin-shielded wire 

M3 Screw Hole

 

Compaction Tube 

Stainless Pipe(SUS304) 

Stainless Pipe(SUS316)

TR-0106  TPE resin-Shielded Sensor 

All dimensions in mm

Cable Length:  0.6m
Thermal-Constant Time

 In the Air:  Approx. 75 Sec.

TR-0206  Stainless Protection Sensor

Cable Length:  0.6m
Thermal-Constant Time

 In the Air:  Approx. 75 Sec.

TR-0306  Stainless Protection Sensor

Cable Length:  0.6m
Thermal-Constant Time

 In agitated water:  Approx. 18 Sec.

 

Only stainless section is water resistant.

TR-0406  Stainless Protection Sensor

Cable Length:  0.6m
Thermal-Constant Time

 In agitated water:  Approx. 20 Sec.

 

Only stainless section is water resistant.

TR-0506  Stainless Protection Sensor

Cable Length:  0.6m
Thermal-Constant Time

 In agitated water:  Approx. 20 Sec.

 

Only stainless section is water resistant.

21

Optional Sensors List

Summary of Contents for TR-71S

Page 1: ...User s Guide Carefully read and fully understand instructions before using this unit 2002 04 16004104060...

Page 2: ...WindowsNT are registered trademarks of Microsoft Corporation USA and are binding in the USA and all other countries Company names and product names are trademarks or registered trademarks of each com...

Page 3: ...ditions with an ambient temperature of 10 to 60 and humidity of 90 RH or less Do not use or store the unit in places such as listed below In places exposed to splashing water In the direct sun light o...

Page 4: ...f time Batteries left in a unit not being used for a long time may leak and cause a malfunction Use the sensor only within the measurable temperature range 40 to 110 For extending the temperature sens...

Page 5: ...temperature The TR 72S is a data logger designed to measure display and record temperature on 1 channel and humidity on its other channel The recorded data from either unit can be downloaded to your...

Page 6: ...be used For details see page 22 Optional Sensors List TR 71S TR 72S Connect the Sensors 10 Left Side View Standard Temperature Humidity Sensor TR 3110 Standard Temperature Sensors TR 0106 Connection c...

Page 7: ...eries and Turn Power On If during recording the switch is turned OFF recording will stop and all recorded data will be stored in the main unit Power Switch ON OFF 11 Caution The replacement period for...

Page 8: ...Recording Conditions settings 13 Battery Life 1 Approximate time till BAT mark appears on the LCD display after installing a new set of batteries is as follows Recording interval of 2 sec or longer A...

Page 9: ...via Main Unit Recording can be started and interval settings made via the buttons on the main unit Only Recording Start Stop and Recording Interval can be controlled by using the main unit buttons Oth...

Page 10: ...can be set via your computer by using the software Thermo Recorder for Windows The following settings can also will be made For more details see the Thermo Recorder for Windows User s Manual or the HE...

Page 11: ...TR 71S TR 72S 2 Open Thermo Recorder for Windows on your computer and make settings such as Start Recording and other Recording Conditions For more details see Thermo Recorder for Windows User s Manu...

Page 12: ...er for Windows User s Manual or the HELP menu of the software Temperature Sensors for TR 7 Series Possible Measurement Range 40 to 110 Sensor Temperature Durability 50 to 115 Measurement Accuracy Aver...

Page 13: ...36 Sec In agitated water Approx 7 Sec TR 1320 Stainless Protection Sensor Cable Length 2m Thermal Constant Time In the Air Approx 12 Sec In agitated water Approx 2 Sec 24 Temperature and Humidity Sen...

Page 14: ...ttery Battery Life Data Back up Interface Communication Speed Dimensions Weight Temp and Humidity Durability of the Unit Standard Sensors Accessories 25 Product Specifications 1 If the unit does not w...

Page 15: ...5652 169 Sasaga Matsumoto City NAGANO 399 0033 JAPAN Phone 81 263 27 2131 Facsimile 81 263 26 4281 Homepage http www tandd co jp english E Mail overseas tandd co jp...