18
Measuring Temperature
1.
Connect the thermocouple wire or thermocouple
extension lead wire to the proper terminals of the
MCR-4TC.
1. Prepare a single wire of Ø0.32mm to Ø0.65mm
(AWG 28-22), or a twisted wire of 0.08mm
2
to
0.32mm
2
(AWG 28-22) with a diameter of 0.12mm
or more. Strip about 10mm from its end.
2. Using a screwdriver or other such tool, while
pressing down on the input terminal button <B> on the top of the
device, insert the wire into the hole <A>.
* Thermocouples have polarity; make sure to
connect in the right way to get a proper
reading.
If the polarity is incorrect, changes in
temperature will be reversed (when the real
temperature increases, the measured
temperature will decrease).
Enlarged View of
Input Terminal
[Top Surface]
A
B
3. When removing the wire, push down on <B> and gently pull the wire
out of the hole <A>.
* MCR-4TC supports the following thermocouple types: K, R, S, T, and J.
* Temperature can only be measured within the heat-durability range of the sensor
being used.
2.
Press <DISPLAY> to switch to the graph display, and view the
measurements in a trend graph.
Measurements
Trend Graph
10mm