14
APM Temp cal
Temp unit(
o
C)
Temp
(
o
C)
POS 0 40 80
Offset +1.0 +0.0 +1.0 -1.0
80
o
C
42
o
C
41
o
C
40
o
C
0
o
C
0
o
C
40
o
C
80
o
C
Actual Value of the Temperature Sensor
1
o
C
Previous curve
Fig. 9 Correction by POS
(3)
Entering Correction Value
(a) Press
Right-Left
key to select
Temp cal
(blinking) from the MENU screen, then press
FUNCTION
key. The Temperature Correction Screen shown in
Fig. 7
will appear.
(b) Press
FUNCTION
key once. (only when the optional software is installed)
(c) Press
Right-Left
key to select correction point (each value is blinking), then change the
offset value by pressing
Up-Down
key. For the next temperature range, press
FUNCTION
key.
(d) Press
RESET
key to return.
2
.
4
External Output Settings
Press
Right-Left
key to select
Ext.out
(blinking) from the MENU screen, then press
FUNCTION
key. The following External Output Setting Screen will appear.
Press
Right-Left
key to select items (each value is blinking), then change the settings.
The explanations of each item are as follows.
Fig. 10 External Output Setting Screen
(1)
RS_232C
Computer
: Data will be transmitted to connected PC after detecting flash point by free run
method. (EXP.FP, FP, etc.) See
2
.
4
.
1
.
Printer
:
Data will be transmitted to connected printer (*1) after detecting flash point.
(EXP.FP, FP, Sample ID, Time, etc.)
APM Ext.out RS_232C
Computer
Inst No. Test No.
95 0037