Parameter Settings
Press
Right-Left
key to select
Para set
(blinking) from the MENU Screen, then press
FUNCTION
key. The following Parameter Setting Screen will appear.
Press
Right-Left
key to select items (each value is blinking), then change the value by pressing
Up-Down
key.
The explanations of each item are as follows.
(1)
I.heat (Initial Heat)
Standard Value is as follows.
Table 2 Initial Heat Values
AC
Source
100V 110V 115V 120V 220V 230V 240V
Standard
Value 26 24 22 20 22 20 18
Too strong an initial heat would cause hunching of temperature rise rate. Too weak an initial
heat could not reach the set rate. Please input an even number (every 2%).
(2)
Cool (Cooling Time)
Cooling time can be change from
00min
(No Cooling) to
99min
(Continuous Cooling).
(3)
First appl (First Application Temperature below EXP.FP)
For APM-7, the temperature of first application of ignition source is changeable. The range is
18
o
C (32
o
F) to 28
o
C (50
o
F) bolow the expected flash point.
Fig. 6 Parameter Setting Screen
APM Para set
First
I.heat Cool appl
22% 10min 22
o
C