The cause for this relatively high temperature is the self 
heating of the bath due to friction-heat of the pump. The 
insulation  of  the  bath  prevents  this  energy  to  be 
transferred into the environment. This causes a raise of 
temperature in the bath up to 25 .. 30°C depending on the 
environmental temperature and used bath fluid. 
  
For  cooling  the  bath  down  to  lower  temperatures,  tap-
water  is  mostly  used.    Also  a  small  circulator  can  be 
connected to the cooling coil like the Tamson TLC10-3, 
or TLC15. 
 

6.4 

Over-temperature safety 

 
In case of electronic failure the possibility exists that the 
heater  element  is  continuously  switched  on.  This  will 
cause  extreme  temperature  raise.  To  prevent  high 
temperatures  the  bath  is  fitted  with  a  mechanical 
temperature  switch  (thermostat).  This  thermostat  will 
switch-off the bath at a preset temperature in the range 
from 50 tot 270°C. 
 
We advise to adjust the mechanical over temperature to 
approxi 25 C above the bath set point.  
 
This  safety  construction  prevents  for  example  oil  to  be 
heated above flash point which definetly will cause fire or 
prevent  evaporation  of  bath  fluid  due  to  high 
temperatures. 
 
The  thermostat  will  automatically  reset  when  the  bath 
temperature drops approximately 10°C below the pre-set 
temperature. To continue normal operation the bath has 
to be switched-off and on again. 
 

6.5 

Adjusting 

The 

Mechanical 

Over-

Temperature Protection Thermostat 

 
• 

Turn the over-temperature protection (8, drawing 

2) to its maximum (turn clockwise towards 50°C), 
• 

Heat the bath to its set point, 

• 

Gently turn the over-temperature protection anti-

clockwise,  until  the  over-temperature  protection  is 
activated, and system switches-off, 
• 

Turn 

the 

over-temperature 

protection 

approximately 25°C higher (turn clockwise), 
Switch  the  bath  off  and  on  again.  The  bath  is  ready  to 
operate safely. 
 
 
 

 

Summary of Contents for TC16

Page 1: ...l sales tamson com Website www tamson com Tamson Instruments bv Page 1 20 tc16 40 58 man docx Rev 2 4 UK 9 21 ISO 9001 2015 NL PRO 238239125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350...

Page 2: ...TION TO THE TC SERIES 8 6 1 GENERAL 8 6 2 PUMP 8 6 3 COOLING 8 6 4 OVER TEMPERATURE SAFETY 9 6 5 ADJUSTING THE MECHANICAL OVER TEMPERATURE PROTECTION THERMOSTAT 9 6 6 FLUID LEVEL 10 6 7 FLUID LEVEL DE...

Page 3: ...damage to the equipment or its accessories caused by incorrect operation Important When performing service maintenance or moving the apparatus always disconnect the apparatus at the main s socket Prop...

Page 4: ...el Avoid damage to the equipment or its accessories through incorrect operation Environment Panel sealing Confirms EN60529 IP65 Environment Temperature 0 tot 35 C Supply enough ventilation Humidity 5...

Page 5: ...ate included in the box as well as ASTM thermometers thermometer holders etc Please see the packing list for details concerning total contents of consignment Before filling the bath remove any remaini...

Page 6: ...ting temperature When viscosity of the bath fluid is too high at room temperature the motor fuse will be activated In such a case the bath fluid can first be heated to set point temperature When the b...

Page 7: ...370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 Do not use demineralized or distilled water This water can lead to corrosion of the bath and moving parts The corrosion can wear out the pump b...

Page 8: ...rough a closed external system The standard pump offers a motor protection small thermal fuse which is placed on the back panel of the motor compartment This protects the motor from excessive loads Wh...

Page 9: ...l cause extreme temperature raise To prevent high temperatures the bath is fitted with a mechanical temperature switch thermostat This thermostat will switch off the bath at a preset temperature in th...

Page 10: ...too low When the bath has been installed it must be filled with an appropriate liquid When working with water the bath should be filled to 1 cm below the lid For oil the bath should be filled to not...

Page 11: ...g on If the bath has been properly filled with fluid it can be switched with the mains switch located on the front panel Choose a working temperature set point 7 2 Control Panel 1 LED Display 2 LED Ov...

Page 12: ...7 Switch for fan only for TC16 with p n 31T0671 31T0861 8 Fan only for TC16 with p n 31T0671 31T0861 7 4 Selecting parameters The front panel layout shows the following four operating keys Press short...

Page 13: ...D values can not be set manually anymore pb Changing the tuning PID parameters Proportional band value Default value is 1 range 1 100 Proportional band equals 1 proportional value Press page key to co...

Page 14: ...ir own heat capacity the use of external cooling and external connected processes circulation or working at different set point temperatures requires new settings of the PID parameters These parameter...

Page 15: ...elow provides fast finding of the PID settings and is refereed to as the method of Ziegler Nichols This procedure is only to be followed when automatic tuning is not functioning This method provides f...

Page 16: ...nt temperature to remove the fluid make sure that the hose to drain and waste container can withstand the temperature of the bath fluid When removing bath fluid do not inhale toxic vapor Always use ap...

Page 17: ...nstruments bv Problems with set point Heater LED is not burning motor is turning and temperature raises above set point Setpoint too near to room temperature Cooling of the bath is needed Temperature...

Page 18: ...error Auto reset after 2 s E6 Watchdog error Auto reset after 2 s 10 SPECIFICATION 10 1 Technical specifications overview Item Unit TC16 TC40 TC58 Ordering code 230V 50 or 60Hz 00T0671 00T0681 00T069...

Page 19: ...safe handling procedures should be provided to handlers and users Alteration of this document is strictly prohibited Except to the extent required by law republication or retransmission of this docum...

Page 20: ...6 42 EC Machinery Directive of the European Parliament and of the Council of 17 May 2006 42 EC 2nd Edition June 2010 p d EN 60204 Machinery Directive and Safety requirements p pi e EN60950 1 Low Volta...

Reviews: