4.5
Fluid Level
Do not operate the bath with low fluid level.
When the fluid level is too low, bath fluid will vaporize
leading to toxic and flamable fumes.
Flamable fumes can lead to fire
Flamable fumes can be ignited bij the not submersed part
of the heating element.
The level indicator(optional) will start to blink (blue light)
when the fluid level is too low.
When the bath has been installed it must be filled
with an appropriate liquid. When working with water
the bath should be filled to 1 cm below the lid. For oil
the bath should be filled to not more than 5 cm below
the lid. Depending on the operating temperature the
liquid level in the bath should be observed and
excessive fluid should be removed.
The liquid level should be maintained between 1 and 3
cm below the lid during normal operation.
The heating element will be damaged when not fully
submerged in the bath fluid. A lower level than 5 cm
below the lid may damage the heaters. A high bath level
can cause overflow and will might also damage the bath
insulation.
Low fluid level
-
Can cause fire when heater get's partially
exposed
-
Will damage the heater