background image

4. Check that the chair operates normally.

3. Connect the power plug with the outlet.

Installation Procedures

1E04K4BE BOOK NO.

1

LANCER Installation Manual

Rubber ring

Base

Base

Foot switches

1. Move the chair to the installation location.

Be sure to ground the chair to prevent accident due to current leakage and to improve the noise-dampening 

performance of the control circuit.

Takara Belmont has no responsibility for any current leakage that results if the above-indicated procedures are 

not performed.

The chair must absolutely be grounded in order to prevent electric shock.

Be sure to store this manual in a secure location together with the operation manual as it describes 

items that must be understood by the customer.

Move the chair by holding the skid.

(Carrying the chair by the upper part is dangerous 

because the base could come off and drop.)

Remove the chair from the skid. Raise the 

chair off the floor by holding the metal plate 

under the foot switch. Move the chair by 

holding the base.

2. Attach the rubber ring to the base.

Determine the installation location. Lift the base off the floor and attach the rubber ring to the rim of the base.

It is easier to attach the rubber ring by lifting up one side of the chair at a time.  Hold the armrest and shoulder 

part of the backrest to balance the force, and tilt the chair body to attach the rubber ring one side at a time.

Do not apply excessive force to the armrest or other parts, or strike it from above when tilting the chair body.

Prepare an available outlet 

with an attached ground terminal.

110 V, 

120V

, 220 V, 230 V, 240 V (7 A or more)

Turn the power on and check that the chair raises/lowers normally, the backrest raises/reclines normally, the 

legrest raises/lowers normally(

AB-HPPN

),

 and the rotation lock/release, preset feature, and auto-return features

operate normally. 

Bang

Summary of Contents for LANCER

Page 1: ...ause the base could come off and drop Remove the chair from the skid Raise the chair off the floor by holding the metal plate under the foot switch Move the chair by holding the base 2 Attach the rubber ring to the base Determine the installation location Lift the base off the floor and attach the rubber ring to the rim of the base It is easier to attach the rubber ring by lifting up one side of t...

Page 2: ...hair from overturning is located on the underside of the step To adjust the adjuster loosen the plating hexagonal nut M10 by using a wrench No 17 to adjust the height of the adjuster and then tighten the plating hexagonal nut M10 by using a wrench No 17 Be sure to securely tighten the plating flange nut M6 to install the step Failure to do so could result in serious injury due to the chair tipping...

Page 3: ...Outer 3 Press and hold the preset switch that you want to set and turn on the power switch Press and hold the preset switch located on either the left or right side of the armrest control panel P1 or P2 and turn on the power switch Keep pressing the preset switch for approximately three seconds Keep pressing until you hear an electronic beep 4 Setting is completed when the electronic beep is emitt...

Page 4: ...o set the legrest rising angle to Setting 1 insert the legrest adjustment pin into the Hole for Setting 1 2 To set the legrest rising angle to Setting 2 insert the legrest adjustment pin into the Hole for Setting 2 3 Do not insert the legrest adjustment pin into any holes if you want the legrest not to rise Carefully store the legrest adjustment pin in order not to lose it Insert the legrest adjus...

Reviews: