Error Messages
MESSAGE
WHAT IT MEANS
WHAT TO DO
E-b
Appears when the batteries
are too low.
Replace the batteries
immediately.
E-U
Appears when a used test
strip is inserted.
Repeat with a new test strip.
E-t
Appears when ambient
temperature is above or
below system operation
range.
System operation range is
10°C to 40°C (50°F to 104°F).
Repeat the test after the meter
and test strip are in the above
temperature range.
E-0,
E-A,
E-C
Problem with the meter.
Repeat the test with a new
test strip. If the meter still does
not work, please contact the
customer service for assistance.
E-F
Appears when test strip is
removed while counting
down, or insufficient blood
volume.
Review the instructions and
repeat test with a new strip.
If the problem persists, please
contact the local customer
service for help.
E-E
Appears when the calibration
data is not correct.
Please contact local customer
service for help.
E-3
Problem with the voice chip.
Please contact local customer
service for help.